Abstract
A repository of 14 gaseous organic compounds at parts-per-billion levels (7-430 ppb) in compressed gas cylinders has been established by the Environmental Protection Agency. This repository was established to provide audit materials for use in conducting performance audits to assess the accuracy and precision of principal organic hazardous constituent (POHC) measurements, especially those performed during hazardous waste trial burn tests. Five- and nine-component mixtures In a balance gas of nitrogen were prepared in aluminum cylinders for use in the performance audit program. The five-component mixtures contain carbon tetrachloride, chloroform, perchloroethylene, vinyl chloride, and benzene. The ninecomponent mixtures contain trichloroethylene, 1,2-dichloroethane, 1,2-dibromoethane, acetonitrile, trichlorofjuoromethane, dichlorodifluoromethane, bromomethane, methyl ethyl ketone and 1,1,1-trichloroethane.
Studies of all 14 gaseous organic compounds were performed to determine the stability of the compounds and the feasibility of using them in performance audits. Results indicate that ail of the 14 compounds were adequately stable to be used as reliable audit materials. Also, the estimation of specific uncertainties associated with the analysis of these audit materials and how those uncertainties are used to assign a total uncertainty to the final analyte concentration for 14 organics Is described.
Subsequent to completion of the stability studies, several performance audits were conducted using the repository gases to assess the accuracy and precision of the volatile organic sampling train (VOST) and bag measurement methods. The audit results indicate that laboratories can use these methods to analyze POHC with accuracy usually to within ± 50 percent of the audit concentrations.