Indirect Determination of Ranitidine Hydrochloride Using a Chloride Ion Selective Electrode

, , , &
Pages 2928-2936 | Received 15 Apr 2009, Accepted 22 Jul 2009, Published online: 02 Nov 2009
 

Abstract

A simple, rapid, and reliable potentiometric method is described for determination of ranitidine hydrochloride in pharmaceutical formulations. Experiments were performed to determine the suitability of using a chloride-ion selective electrode for the indirect estimation of ranitidine hydrochloride. The precision and accuracy of the potentiometric method were determined. We found that they not differ significantly. The advantage of the proposed method is the fact that it can be applied without filtration of the drug sample before analysis.

Notes

∗Mean of five independent measurements.

a Confidence level was calculated for t (Student test) = 3.18; n 1 (nr. of potentiometric measurements) = 4 and P (Probability) = 95%.

b Confidence level was calculated for t = 4.3; n 2 (no. of spectrophotometric measurements) = 3 and P = 95%.

# F tabulated  = 19.20 for n 1 = 4; n 2 = 3 and P = 95%.

a Mean of three measurements ± standard deviation (S).

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.