9
Views
5
CrossRef citations to date
0
Altmetric
Electrodes/Biosensors

Potentiometric Enzyme-Amplified Flow Injection Analysis Detection System: Behavior of Free and Liposome-Released Peroxidase

&
Pages 1107-1124 | Received 18 Jan 1989, Accepted 20 Jan 1989, Published online: 23 Oct 2006
 

Abstract

A flow injection analysis (FIA) system was designed for the determination of free and liposome-released peroxidase activity. the inclusion of an extra washing channel in the FIA system provided a solution to problems of electrode fouling and slow recovery of response. the addition of Triton X-100 to acetate buffer resulted in an effective and reliable wash solution, and improved the electrode recovery rate by a factor of three. the sensitivity also increases by about 20% for the enzyme with 3 mU activity by using the wash solution containing Triton X-100.