A Fluorimetric Method for the Determination of Cyanide with Fluorescein and IodineFootnote

, &
Pages 2797-2803 | Received 17 May 1994, Accepted 12 Jul 1994, Published online: 22 Aug 2006
 

Abstract

A rapid, simple and sensitive fluorimetric method has been developed for the determination of cyanide with fluorescein as fluorogenic reagent (λex = 494 nm, λem = 514 nm) at pH 6.0–7.0. A linear calibration curve was obtained in the range 0.004–2.0 μg CN/25 ml. The detection limit is 0.004 μg CN-/25 ml. The method was successfully applied to the determination of cyanide in waste water.