A Neutral Carrier Membrane Calcium-Sensitive Electrode with Wide Working pH Range

, , &
Pages 1637-1646 | Received 06 Dec 1993, Accepted 14 Mar 1994, Published online: 20 Aug 2006
 

Abstract

A poly(vinyl chloride) (PVC) membrane calcium-sensitive electrode based on a new type of synthetic carrier 1,7-di[2-(1-phenylazo)naphtyl]-1,4,7-trioxaheptan is introduced. This electrode exhibited Nernstian response over the concentration range 1.0×10−1 - 3.0×10−8 mol/L Ca2+ with a slope of 28.8 mV/pCa in calcium ion buffer solutions. The detection limit for calcium ion was 1.0X10−8 mol/L. The response time varied from 10 to 30 seconds. The working pH range for the electrode was 3.5 - 12.3 when the concentration of Ca2+ was 10−3 mol/L. The lifetime of the electrode was proved to exceed 8 months. It was applied successfully in the determination of calcium ion in river water, mineral water and cosmetic. A method for the determination of total calcium in urine was developed.