Determination of Pimethixene in Plasma and Urine by High Performance Liquid Chromatography

&
Pages 775-784 | Received 21 Dec 1994, Accepted 20 Nov 1995, Published online: 22 Aug 2006
 

Abstract

A high performance liquid chromatographic (HPLC) assay was developed for the determination of pimethixene in plasma and urine. Pimethixene was extracted by a rapid and simple liquid-liquid procedure using n-pentane:isopropyl alcohol (50:1), and a reversed phase HPLC separation technique with UV spectrophotometric detection at 254 nm. Determination was possible for pimethixene in the concentration range 53–530 ng ml with the limit of detection at 20 ng/ml for both plasma and urine. The overall recoveries of pimethixene added to plasma and urine samples by an external standard method were 95.6 and 98.5%, with coefficients of variation of 0.9993 and 0.9997, respectively.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.