Abstract
In 2009 the Alcohol Test Committee of the Canadian Society of Forensic Science changed the calibration check interval of Approved Screening Devices from every 2 weeks to every 31 days. Data from 612 calibration checks of a group of Alco-Sensor IV DWF screening devices were analyzed using a contingency table and x2 analysis as well as a significance test for a difference in proportions. There were no statistically significant differences between the length of time of the calibration interval and the frequency of a device requiring recalibration. This analysis shows the Alco-Sensor IV DWF is no more likely to require calibration up to and including a 5 week interval when compared to a 2 week interval, supporting the Alcohol Test Committee decision.
Résumé
En 2009, le Comité des analyses d'alcool de la Société canadienne des sciences judiciaires a modifié l'intervalle de temps requis entre les vérifications de l'étalonnage des appareils de détection approuvés de 2 semaines à 31 jours. Les données provenant de 612 vérifications de l'étalonnage d'un groupe d'appareils de détection du type Alco-Sensor IV DWF ont été analysées à l'aide d'un tableau de contingence et d'analyse x2 ainsi qu'à l'aide d'un test de « significance » pour les différences de proportions. Aucune différence statistiquement significative n'a été trouvée entre l'intervalle de temps entre les vérifications de l'étalonnage et la fréquence avec laquelle un appareil devait être ré-étalonné. Cette analyse démontre que l'étalonnage de l'Alco-Sensor IV DWF n'a pas à être refait durant une période de 5 semaines supportant donc la décision du Comité des analyses d'alcool.