Performance of automated urine analyzers using flow cytometric and digital image-based technology in routine urinalysis

ORCID Icon, ORCID Icon, ORCID Icon & ORCID Icon
Pages 468-474 | Received 19 Dec 2018, Accepted 19 Aug 2019, Published online: 28 Aug 2019
 

Abstract

The purpose of this study was to evaluate the analytical performances of Sysmex UF-5000 and Dirui FUS-200 and to compare the results with manual microscopy and between each other.

Two hundred fifty urine samples were analyzed for evaluation. Mid-stream specimens were studied sequentially using Dirui FUS-200 and Sysmex UF-5000, and also with manual microscopy within one hour. The physical and chemical components of urinalysis, and sediment results were investigated.

The precision results of the FUS-200 and UF-5000 for WBCs, RBCs, and ECs were acceptable. The both analyzers demonstrated good linearity (r > 0.97), with no carry-over. The comparisons of FUS-200 and UF-5000 with manual microscopy for RBCs, WBCs, and ECs on 250 samples exhibited good agreement with little bias (R > 0.780). Only, the moderate agreements were obtained for calcium oxalate for both analyzers (R = 0.512, and 0.648, respectively). The sensitivities of the FUS-200 and UF-5000 were 75.8% and 86.8%, with specificities of 92.3% and 87.8% for WBCs, for RBCs the sensitivities were 91.1%, and 84.4% with specificities of 82.2%, and 89.6% for both analyzers. Kappa values of the UF-5000 were higher than FUS-200 for WBCs, RBCs, ECs, and calcium oxalate.

The FUS-200 and UF-5000 urine analyzers, are both accurate, very precise systems and can be safely used in clinical laboratories. However, due to the technological characteristics of the UF-5000 analyzer, its positive impacts on the morphologic recognition and enumeration of RBCs and WBCs should be taken into account, particularly in university hospital laboratories with high patient volumes.

Acknowledgment

Authors thank to Sysmex Turkey for providing UF-5000 urine analyzer and all the technical assistance.

Disclosure statement

No potential conflict of interest was reported by the authors.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.