Development of a designated comparison method for alkaline phosphatase measurements and its application to evaluating routine methods

, , , , , , & show all
Pages 218-224 | Received 05 Nov 2020, Accepted 07 Mar 2021, Published online: 23 Mar 2021
 

Abstract

The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) published the reference measurement procedure (RMP) for ALP measurement in 2011. However, the RMP is of high requirements for laboratories, complicated, time-consuming and high cost of reagents. Many manufacturers do not trace results to the higher procedure. And there is currently no designated comparison method (DCM) for ALP measurement. Thus, the standardization of ALP measurement is hindered. Automatic biochemical analyzers are easy to operate and widely used in clinical laboratories. Therefore, according to the RMP, establishing a DCM based on an automatic biochemical analyzer will be a practical way to establish traceability to the accuracy base and promote the standardization of ALP measurement. On the basis of conforming to the RMP recommended by IFCC as far as possible, the DCM was established based on a Thermo Indiko automatic biochemical analyzer. Performances of the method were validated. The DCM repeatability and within laboratory imprecision was <1% and <2.5%, respectively. For evaluation of trueness, the biases were within the equivalent limits. Measurement procedure comparisons and biases estimation were carried out between the DCM, the RMP, and the six routine methods using a panel of 40 individual human serum samples. The comparisons between the DCM and the RMP gave satisfying results. Compared with the DCM, the relative biases of some routine methods failed to meet the bias limit derived from biological variation.

Disclosure statement

No potential conflict of interest was reported by the author(s).

Additional information

Funding

This work was supported by The Leaping Development Project of Beijing Biomedical Industry under Grant No. Z161100001816043.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.