Abbott D-dimer assay: analytical performance and diagnostic accuracy in management of venous thromboembolism

, , &
Pages 53-61 | Received 01 Aug 2023, Accepted 10 Feb 2024, Published online: 16 Feb 2024
 

Abstract

This study aimed to assess analytical characteristics and diagnostic accuracy in management of venous thromboembolism (VTE) in the Emergency Department (ED) of the Abbott D-dimer assay applied on the Alinity c clinical chemistry analyzer (Abbott Laboratories, Chicago, IL) compared to the INNOVANCE D-dimer assay (Siemens Healthineers, Marburg, Germany). Precision was determined at three concentration levels following the CLSI EP15-A3 protocol. Method comparison and diagnostic accuracy were assessed using samples obtained from 85 patients who were referred for diagnostic imaging and D-dimer testing due to clinically suspected VTE. Within-run coefficients of variation (CVs) were 3.0%, 0.5% and 0.5% at D-dimer concentrations of 0.54, 1.42 and 2.68 mg/L FEU, while respective between-run CVs were 2.0%, 3.4% and 2.7%, hence fulfilling the desirable biological variation criteria for imprecision (<12.6%). Passing–Bablok regression analysis yielded a small proportional difference between the two compared assays (y = 1.09 (95% confidence interval (CI): 1.01–1.18) x + 0.09 (95%CI: –0.09 to 0.16)), while Bland–Altman analysis showed significant negative absolute (–0.6 mg/L FEU, 95%CI: –0.9 to −0.3) and relative mean bias (–14.1%, 95%CI: –20.3 to −7.9). Spearman’s ρ was 0.979 (95%CI: 0.967–0.986). Inter-assay agreement relative to the cut-off was 92% (kappa coefficient = 0.547 (95%CI: 0.255–0.839)). Diagnostic sensitivity, specificity, positive and negative predictive values of the Abbott assay were 100%, 9.2%, 25.3% and 100%, respectively, compared to the following data for the INNOVANCE assay: 95.0%, 15.4%, 25.7% and 90.9%. Abbott D-dimer assay has shown excellent analytical precision, high comparability with the INNOVANCE D-dimer and high NPV at manufacturer’s cut-off.

Disclosure statement

No potential conflict of interest was reported by the author(s).

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.