Determination of coagulation factors (II-VII-X) and kaolin partial thromboplastin time by the LKB 8600 Reaction Rate Analyser

&
Pages 15-20 | Received 17 Nov 1976, Accepted 08 Jan 1977, Published online: 28 Aug 2009
 

Abstract

Testing of a modified Reaction Rate Analyser 8600 in the automatic determination of coagulation time was accomplished with the assays Thrombotest®, Normotest® and Kaolin Partial Thromboplastin Time. Variations in extinctions, from the adding of start reagent until coagulation, were recorded. The coagulation activities, found by the automatic method for the assays Thrombotest® and Normotest®, were compared with a manual method. Thrombotest® showed a good correlation: 0.30 > P > 0.25 (paired t test). Normotest® showed significantly different values by the methods: P<0.001 (paired t test). Kaolin Partial Thromboplastin Time (Activated Partial Thromboplastin Time with kaolin as activator) gave significantly lower values— the difference was 3 sec—compared with a manual method, and the precision was considerably improved.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.