Comparison of intraocular pressure in New Zealand White rabbits measured using rebound and applanation tonometers and four different methods of physical restraint

ORCID Icon, ORCID Icon, ORCID Icon, ORCID Icon, ORCID Icon, ORCID Icon, ORCID Icon & ORCID Icon show all
Pages 251-258 | Received 10 Feb 2023, Accepted 31 May 2023, Published online: 26 Jun 2023
 

ABSTRACT

Aims

To compare intraocular pressure (IOP) measurements obtained in rabbits using rebound (TV) and applanation (TPV) tonometers with four different methods of physical restraint.

Methods

A total of 20 New Zealand White rabbits (40 eyes) were included in this study. IOP readings were obtained from both eyes using the two different tonometers. The rabbits were placed on a table and restrained by wrapping in a cloth (Method I), by scruffing with rear support (Method II), by wrapping in a cloth and cupped in the hands (Method III), or by a box restrainer (Method IV).

Results

The mean IOP measurement obtained by TPV was higher than that obtained with the TV for all handling methods. Mean differences (TV−TPV, in mmHg) in IOP were −5.3 (95% Cl = −6.5 to −4.1) for Method 1, −4.7 (95% Cl = −6.2 to −3.29) for Method II, −4.9 (95% Cl = −6.2 to −3.7) for Method III and −7.6 (95% Cl = −9.2 to −5.9) for Method IV. Using the TV tonometer, mean IOP for Method IV was higher than for Method I (mean difference 2.1 (95% Cl = 1.1–3.1)), whereas using the TPV tonometer, mean IOP for Method IV was significantly higher than Method I, II, and III (mean differences: 4.4 (95% Cl = 2.6–5.9), 3.7 (95% Cl = 2–5.3) and 3.8 (95% Cl = 2–5.4), respectively). According to Bland-Altman plots, IOP readings for TPV tended to be higher than those for TV with all handling methods, but with a lack of agreement. The mean difference and 95% limits of agreement for the differences between TV and TPV were −5.4 mmHg (−12.5–1.9 mmHg), −4.7 mmHg (−12.9–3.5 mmHg), −4.9 mmHg (−12–2.2 mmHg), and −7.5 mmHg (−17.4–2.3 mmHg), with Methods I, II, III, and IV, respectively. Comparing TV and TPV, only 7.5%, 12.5%, 27.5%, and 15% of IOP measurements from 20 rabbits were within the range considered clinically acceptable for IOP (± 2 mmHg) for Method I, II, III, and IV, respectively.

Conclusion and clinical relevance

In conclusion, the physical restraint method should be recorded when IOP is measured in rabbits, and TV and TPV tonometers cannot be used interchangeably (high bias and low proportion of measurements within ± 2 mmHg).

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.