A comparison of wind noise in four hearing instruments

, , &
Pages 171-174 | Published online: 12 Oct 2009
 

Abstract

Wind noise was measured in four behind-the-ear hearing instruments with different microphone openings. A silent airflow of 7 m/s was directed toward the ear of a Kemar head and the resulting wind noise was measured. The amplification was set to an insertion gain of 35 dB at 1.6 kHz. The wind noise amplitude at the position of the drum ranged from 84 to 97 dB(A). The hearing instrument with a partially covered microphone entrance proved best, whereas one with an open microphone entrance had the poorest performance. A reduction in wind noise of 6-17 dB could be achieved in all hearing instruments by using a simple windscreen made of Styrofoam. The windscreen affected the frequency response by less than 2 dB. By subtracting the insertion gain from the wind noise, an equivalent wind noise could be presented as a function of frequency. A considerable difference was found between the wind noise sensitivity in different hearing instruments. All could be improved by a windscreen without adversely affecting the frequency response.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.