Theoretical investigation of measurement procedures for the quality assurance of superficial hyperthermia applicators

, &
Pages 416-425 | Published online: 09 Jul 2009
 

Abstract

This work presents the results obtained from simple numerical models concerning the measurement uncertainty with thermographic techniques used for the evaluation of superficial hyperthermia applicators. Based upon the calculations performed, it is shown that, when using a thermographic technique to measure the SAR distribution of an applicator, heating times from 60-120s and measuring times of 10s are acceptable for an accurate assessment of the half-width at half power (HWHP) of an applicator (error less than 2%) with an expected HWHP larger than 2.5cm. Only when the HWHP is expected to be less than 2.5cm does the heating time need to be adapted to obtain an accuracy of 2% or better. For the assessment of the maximum SAR, the situation is worse. Even with a careful experimental design, it is difficult to measure the maximum SAR with an error less than 7%.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.