Characterization of the SAR-distribution of the Sigma-60 applicator for regional hyperthermia using a Schottky diode sheet

, , , , &
Pages 642-654 | Received 04 Feb 2003, Accepted 01 May 2003, Published online: 09 Jul 2009
 

Abstract

Introduction: Characterization of the performance of an hyperthermia applicator by phantom experiments is an essential aspect of quality assurance in hyperthermia. The objective of this study was to quantitatively characterize the energy distribution of the Sigma-60 applicator of the BSD2000 phased array system operated within the normal frequency range of 70–120 MHz. Additionally, the accuracy of the flexible Schottky diode sheet to measure E-field distributions was assessed.

Material and methods: The flexible Schottky diode sheet (SDS) consists of 64 diodes mounted on a flexible 125 µm thick polyester foil. The diodes are connected through high resistive wires to the electronic readout system. With the SDS E-field distributions were measured with a resolution of 2.5 × 2.5 cm in a cylindrical phantom, diameter of 26 cm and filled with saline water (2 g/l). The phantom was positioned symmetrically in the Sigma-60 applicator. RF-power was applied to the 4-channel applicator with increasing steps from 25 W to a total output of 400 W.

Results: The complete system to measure the E-field distribution worked fine and reliably within the Sigma-60 applicator. The E-field distributions measured showed that the longitudinal length of the E-field distribution is more or less constant, e.g. 21–19 cm, over the frequency range of 70–120 MHz, respectively. As expected, the radial E-field distributions show a better focusing towards the centre of the phantom for higher frequencies, e.g. from 15.3–8.7 cm diameter for 70–120 MHz, respectively. The focusing target could be moved accurately from the left to the right side of the phantom. Further it was found that the sensitivity variation of nine diodes located at the centre of the phantom was very small, e.g. <3% over the whole frequency range.

Conclusion: The SAR distributions of the Sigma-60 applicator are in good agreement with theoretically expected values. The flexible Schottky diode sheet proves to be an excellent tool to make accurate, quantitative measurements of E-field distributions at low (25 W) and medium (400 W) power levels. An important feature of the SDS is that it enables one to significantly improve quantitative quality assurance procedures and to start quantitative comparisons of the performance of the different deep hyperthermia systems used by the various hyperthermia groups.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.