A feasibility study of interstitial hyperthermia plus external beam radiotherapy in glioblastoma multiforme using the multi electrode current source (MECS) system

, , , , , & show all
Pages 451-463 | Received 24 Sep 2003, Accepted 29 Dec 2003, Published online: 09 Jul 2009
 

Abstract

Purpose: Thermoradiotherapy has been shown in several randomized trials to increase local control compared to radiotherapy alone. The first randomized study of interstitial hyperthermia in glioblastoma multiforme showed a survival benefit for hyperthermia, though small. Improvement of the heating technique could lead to improved results. The purpose of this feasibility study is to present the clinical and thermal data of application of an improved interstitial hyperthermia system.

Methods and materials: Six patients with a glioblastoma multiforme were treated with interstitial hyperthermia using the Multi Electrode Current Source Interstitial Hyperthermia (MECS-IHT) system. The MECS-IHT system has the capability of spatial monitoring of temperature and individually steering of heating electrodes. Three sessions were given aiming at a steady state temperature of 42°C for 1 h, with an interval of 3–4 days, during an external irradiation scheme of 60 Gy in 6 weeks. Hyperthermia was delivered with a mean of 10 catheters, 18 heating electrodes and 38 thermal probes per patient.

Results: Sub-optimal temperatures were encountered in the first two patients leading to adjustments in technique thereafter with subsequent improvement of thermal data. With a catheter spacing of 11–12 mm, measurements yielded a mean T90, T50 and T10 of 39.9, 43.7 and 45.2°C, respectively, over three sessions in the last patient. The power per electrode to reach this temperature distribution varied from 25–100% of full power in each of the last four patients. Thermal data were reproducible over the three sessions. Acute toxicity was minimal.

Conclusions: Despite the spatial steering capabilities of the MECS-IHT system, a large temperature heterogeneity was encountered. The heterogeneity was the reason to limit the catheter spacing to 11–12 mm, thus making only small tumour volumes feasible for interstitial heating.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.