Methods and potentials of magnetic resonance imaging for monitoring radiofrequency hyperthermia in a hybrid system

, , , , , , & show all
Pages 497-513 | Received 08 Nov 2004, Accepted 20 Jan 2005, Published online: 09 Jul 2009
 

Abstract

Introduction: Non-invasive thermometry (NIT) is a valuable and probably indispensable tool for further development of radiofrequency (RF) hyperthermia. A hybridization of an MRI scanner with a hyperthermia system is necessary for a real-time NIT. The selection of the best thermographic method is difficult, because many parameters and attributes have to be considered.

Methods: In the hybrid system (Siemens Symphony/BSD-2000-3D) the standard methods for NIT were tested such as T1, diffusion (ADC: apparent diffusion coefficient) and proton-resonance-frequency shift (PFS) method. A series of three-dimensional datasets was acquired with different gradient-echo sequences, diffusion-weighted EPI spin-echo sequences and calculated MR-temperatures in the software platform AMIRA-HyperPlan. In particular for the PFS-method, corrective methods were developed and tested with respect to drift and other disturbances. Experiments were performed in phantoms and the results compared with direct temperature measurements. Then the procedures were transferred to clinical applications in patients with larger tumours of the lower extremity or the pelvis.

Results: Heating experiments and MR-thermography in a homogeneous cylindrical phantom give an excellent survey over the potentials of the methods. Under clinical conditions all these methods have difficulties due to motion, physiological changes, inhomogeneous composition and susceptibility variations in human tissues. The PFS-method is most stable in patients yielding reasonable MR temperature distributions and time curves for pelvic and lower extremity tumours over realistic treatment times of 60–90 min. Pooled data exist for rectal tumour recurrencies and soft tissue sarcomas. The fat tissue can be used for drift correction in these patients. T1 and diffusion-dependent methods appear less suitable for these patients. The standard methods have different sensitivities with respect to the various error sources. The advantages and pitfalls of every method are discussed with respect to the literature and illustrated by the phantom and patient measurements.

Conclusions: MR-controlled RF hyperthermia in a hybrid system is well established in phantoms and already feasible for patients in the pelvic and lower extremity region. Under optimal conditions the temperature accuracy might be in the range of 0.5°C. However a variety of developments, especially sequences and post-processing modules, are still required for the clinical routine.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.