Evaluation of acoustic-thermal simulations of in vivo magnetic resonance guided focused ultrasound ablative therapy

ORCID Icon, , , , ORCID Icon, ORCID Icon & ORCID Icon show all
Article: 2301489 | Received 13 Sep 2023, Accepted 28 Dec 2023, Published online: 17 Jan 2024
 

Abstract

Purpose

To evaluate numerical simulations of focused ultrasound (FUS) with a rabbit model, comparing simulated heating characteristics with magnetic resonance temperature imaging (MRTI) data collected during in vivo treatment.

Methods

A rabbit model was treated with FUS sonications in the biceps femoris with 3D MRTI collected. Acoustic and thermal properties of the rabbit muscle were determined experimentally. Numerical models of the rabbits were created, and tissue-type-specific properties were assigned. FUS simulations were performed using both the hybrid angular spectrum (HAS) method and k-Wave. Simulated power deposition patterns were converted to temperature maps using a Pennes’ bioheat equation-based thermal solver. Agreement of pressure between the simulation techniques and temperature between the simulation and experimental heating was evaluated. Contributions of scattering and absorption attenuation were considered.

Results

Simulated peak pressures derived using the HAS method exceeded the simulated peak pressures from k-Wave by 1.6 ± 2.7%. The location and FWHM of the peak pressure calculated from HAS and k-Wave showed good agreement. When muscle acoustic absorption value in the simulations was adjusted to approximately 54% of the measured attenuation, the average root-mean-squared error between simulated and experimental spatial-average temperature profiles was 0.046 ± 0.019 °C/W. Mean distance between simulated and experimental COTMs was 3.25 ± 1.37 mm. Transverse FWHMs of simulated sonications were smaller than in in vivo sonications. Longitudinal FWHMs were similar.

Conclusions

Presented results demonstrate agreement between HAS and k-Wave simulations and that FUS simulations can accurately predict focal position and heating for in vivo applications in soft tissue.

Acknowledgments

Mark H. Huntsman Endowed Chair. The authors appreciate help from Dr. Rock J Hadley and Robb Merrill for designing and building the rabbit setup and RF coil used in this study.

Disclosure statement

No potential conflict of interest was reported by the author(s).

Data availability

The data that support the findings of this study are available from the corresponding author, AP, upon reasonable request.

Additional information

Funding

NIH grants R37CA224141, R01EB028316, S10OD018482, and R21EB033117.