Central Corneal Edema with Scleral-Lens Wear

, , &
Pages 1305-1315 | Received 21 Mar 2018, Accepted 22 Jun 2018, Published online: 06 Aug 2018
 

ABSTRACT

Purpose: To evaluate the safety of scleral-lens designs, we model and clinically assess central corneal edema induced by scleral-lens wear for healthy subjects.

Materials and Methods: Central corneal swelling during scleral-lens wear is measured using optical coherence tomography (OCT). Transport resistances are modeled for oxygen diffusion through the scleral lens and post-lens tear-film (PoLTF), and into the cornea. Oxygen deficiency in the cornea activates anaerobic metabolic reactions that induce corneal edema. Oxygen permeability, carbon-dioxide permeability, settled-lens PoLTF thickness, and scleral-lens thickness are varied in the calculations to mimic different lens fits.

Results: Transport modeling predicts that for open eyes, increasing PoLTF thickness from 50 to 400 µm increases central corneal swelling by approximately 1–1.5% when oxygen transmissibility (Dk/L) is greater than 10 hBarrer/cm (i.e., hectoBarrer/cm). Although swelling is larger for oxygen Dk/L < 10 hBarrer/cm, PoLTF thickness has minimal impact in this range. For open eye, oxygen transmissibility of the lens plays a significant role in corneal edema, but is negligible when oxygen Dk/L is > 40 hBarrer/cm. For closed eye, central corneal swelling is greater than 5% for an oxygen Dk/L range of 0–100 hBarrer/cm with typical lens-fitting parameters. For carbon-dioxide transmissibilities increasing from 50 to 250 hBarrer/cm and with a fixed oxygen Dk/L of 25 hBarrer/cm, calculated swelling diminishes by an additional 0.5%. Comparison of model calculations to clinical-swelling data is within the error range of the clinical measurements.

Conclusions: Oxygen/metabolite transport calculations for open-eye scleral-lens wear show that typical PoLTF thicknesses fitted by clinicians (i.e., PoLTF thicknesses < 400 µm) with modern scleral lenses (i.e., oxygen Dk/L > 25 hBarrer/cm) produce corneal swelling of less than 2% in agreement with experiment. Therefore, scleral lenses prescribed today evoke less than physiological hypoxic swelling (i.e., less than 4%) for healthy corneas during open-eye. Closed-eye wear, however, appears clinically unsafe.

Declaration of interest

The authors report no conflicts of interest.

Additional information

Funding

Financial support from Clinical Research Center Unrestricted Fund (MCL) and Lagado Corporation (MCL) is acknowledged.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.