Ocular Cyclorotation and Corneal Axial Misalignment in Femtosecond Laser-Assisted Cataract Surgery

ORCID Icon, , , , , , & show all
Pages 1313-1318 | Received 30 May 2019, Accepted 28 Jun 2019, Published online: 11 Jul 2019
 

ABSTRACT

Purpose: To explore ocular cyclorotation and the source of corneal axial misalignment during femtosecond laser-assisted cataract surgery (FLACS).

Methods: Forty-five sequential patients (50 eyes) who had undergone FLACS (LenSx Laser System, Alcon Inc) were recruited. We took screenshots from videos of FLACS to analyze ocular cyclorotation and the real angle between primary incision and secondary incision (RAPS). In addition, crystalline lens tilt and theoretic angle between the primary and secondary incisions (TAPS) was also calculated.

Results: The mean absolute value of ocular cyclorotation was 8.03 ± 4.48 degrees (0–19.1 degrees). The crystalline lens tilt was 3.30 ± 1.44 degrees (0.93–6.44 degrees). And the mean preoperative uncorrected visual acuity was 0.89 ± 0.50 LogMAR units. Pearson bivariate correlation analysis showed significant positive correlation between ocular cyclorotation with crystalline lens tilt (r = 0.37, p = .008), and ocular cyclorotation negatively correlated with axial length (r = −0.29, p = .038). In addition, the TAPS was 89.78 ± 1.45 degrees, and the RAPS was 85.68 ± 2.04 degrees. The angle error was 4.11 ± 1.28 degrees (p<0.001).

Conclusions: Ocular cyclorotation commonly occurred during FLACS. In addition, increased axial length was associated with less ocular cyclorotation and increased crystalline lens tilt was related to more ocular cyclorotation. Importantly, machinery systemic errors during corneal astigmatism correction by arcuate incision in FLACS should be taken into consideration.

Declaration of interest

No potential conflict of interest was reported by the authors.

Supplementary material

Supplemental data for this article can be accessed here.

Additional information

Funding

The current study was supported by the National Natural Science Foundation of China [grant nos. 81700817].

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.