Computational and Experimental Analysis of an In Vitro Microfluidic Experimental Setup on Testing Molteno, Ahmed Valve and Ex-Press Implants and Their Critical Comparisons

, , &
Pages 69-78 | Received 16 Sep 2020, Accepted 24 Jun 2021, Published online: 21 Oct 2021
 

ABSTRACT

Purpose/Aim of the study

Researchers have investigated glaucoma drainage devices in various in vitro experimental setups to overcome their disadvantages. In this study, it is aimed to perform in vitro analyses of glaucoma drainage devices that are actively utilized for glaucoma treatment.

Materials and Methods

A new in vitro experimental setup is constructed and the first outcomes of numerical simulations and experimental trials are shared. Firstly, an in vitro experimental setup is designed and simulated in ANSYS Fluent, then assembled/fabricated using microfluidic equipment. Secondly, computational fluid dynamics results are expressed as pressure losses through an Ahmed Valve implant, a Molteno drainage device, an Ex-Press implant for physiological flow rates.

Results

In the scope of this study, Ahmed Valve (valved), Molteno (non-valved), Ex-Press (orbital shunt) implants are examined numerically using computational fluid dynamics tools. Results are compared with in vitro studies of the proposed experimental setup. Poiseuille and Reynolds numbers versus pressure drop characteristics of tested glaucoma drainage devices are also obtained using in vitro microfluidic experimental setup. In the range of 1.6–2.5 µl/min, Ahmed Valve implant created an active and effective pressure drop of 5.6 to 12 mmHg. Ahmed Valve and Molteno showed similar characteristics in terms of Poiseuille and Reynolds numbers variations with pressure drop across implants. Ex-Press, opposing Molteno and Ahmed Valve gave a reciprocal correlation between pressure drop and Po.

Conclusions

High resemblances of pressure drops between computational fluid dynamics results and in vitro microfluidic experimental results proved that the setup will be a better choice compared to syringe pump type setups for testing different glaucoma drainage devices at the same conditions. In conclusion, a new glaucoma drainage device should be equipped with a slimmer but larger end plate and a valve mechanism similar to Ahmed Valve in the perspective of intraocular pressure drop performance.

Disclosure statement

No potential conflict of interest was reported by the author(s).

Additional information

Funding

This work was supported by the Scientific and Technological Research Council of Turkey-TÜBİTAK 3001 Project under Grant No. [117M971].

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.