A bipolar electrospray source of singly charged salt clusters of precisely controlled composition

&
Pages 778-786 | Received 21 Oct 2016, Accepted 27 Feb 2017, Published online: 05 Apr 2017

ABSTRACT

The few clusters [BnA+n+1]+ (n = 0,1) with resolvable mobilities formed in electrosprays of large salts have been used for nanoparticle instrument testing and calibration at sizes smaller than 2 nm. Extensions of this modest size range by charge reduction with uncontrolled gas phase ions has resulted in impure singly charged clusters. Here, we combine two oppositely charged electrosprays of solutions of the same salt BA+, including: (CnH2n+1)4N+Br (n = 4,7,12,16), the large phosphonium cation (C6H13)3(C16H33)P+ paired with the anions Im [(CF3SO2)2N] or FAP [(C2F5)3PF3], and the asymmetric pair [1-methyl-3-pentylimidazolium+FAP]. Both polarities are simultaneously produced by this source in comparable abundances, primarily as singly charged A+nBn±1, with tiny contributions from higher charge states. Some but not all of these clusters produce narrow mobility peaks typical of pure ions, even beyond n = 43. Excellent independent stable control of the positive and the negative sprays brought very close to each other is achieved by isolating them electrostatically with a symmetrically interposed metallic screen. Two nanoDMAs covering the size range up to 30 nm (Halfmini and Herrmann DMAs, with classification lengths of 2 and 10 cm) are characterized with these standards, revealing resolving powers considerably higher than previously seen with unipolar electrospray sources. The bipolar source of pure and chemically homogeneous clusters described permits studying size and charge effects in a variety of aerosol instruments in the 1–4 nm size range.

Copyright © 2017 American Association for Aerosol Research

EDITOR :

1. Introduction

Monodisperse particle standards have been widely used in aerosol science, particularly in the evaluation of instruments, including differential mobility analyzers (DMA), impactors, condensation particle counters (CPC), etc. The so-called monodisperse aerosols, traditionally based on polystyrene spheres or size selection with a DMA, have always had a finite size breadth. Nanometer size standards were traditionally of poor monodispersity, until the development of electrospray (ES) ionization (Fenn et al. Citation1989) provided them in essentially pure form. Since then, a variety of strictly monodisperse size or mobility standards have become available, permitting unprecedented studies (Winkler et al. Citation2012). Molecular mobility standards available by conventional electrospraying, however, have been limited to small sizes (1.7 nm) by the fact that ES produces a complex mixture of clusters appearing in many charge states. As a result, most clusters in this mix cannot be resolved from each other by mobility alone, with the exception of the smallest ones that are singly charged and have peculiarly large mobilities (for some exceptions see Ku and Fernandez de la Mora Citation2004, Citation2009). Charge-reduction of multiply charged electrosprayed clusters has permitted a considerable widening of this range (Kaufman et al. Citation1996; Gamero-Castaño and Fernandez de la Mora Citation2000a,Citationb,Citationc; Kaufman and Dorman Citation2008; You et al. Citation2014). The approach, however, reduces the original charge with a mixture of counterions produced by a source of ionizing radiation, and results in approximately monodisperse particles rather than rigorously pure species (Gamero and Fernandez de la Mora Citation2000a,Citationb,Citationc,Citationd; Figures 1b, 4c, and 1 and Figures 3, 4, and 9, respectively). Tandem DMA selection of multiply charged clusters that lose one charge between the two DMAs has increased up to 3.5 nm the available size range (Attoui et al. Citation2013). However, this method requires two high-resolution DMAs, one of which needs to have a high transmission available only in planar DMAs. Kangasluoma (Citation2015) has been especially creative of alternative sources for cluster standards (see also Kangasluoma et al. Citation2016a,Citationb) at sizes well beyond those produced by electrospraying salts. Some of these studies have achieved an improved control of the charge reduction process by selecting the counter-ion. This can be performed to a certain degree by exploiting the specificity of certain chemical ionization processes, for instance by resonant laser ionization (Kane et al. Citation1995; Kane and El-Shall Citation1996) or by selectively ionizing a species with high ion or electron affinity (Reid et al. Citation2003). This approach has even been used successfully for the much harder task of producing neutral clusters of selected composition (Steiner et al. Citation2017). Still, even if an electrospray of BnA+n+z is charge-reduced by a pure ion C (say NO3, which may be selectively created by chemical ionization; Jiang et al. Citation2011), mixed species of the form BnA+mCm-n±1 result. The double degree of freedom m, n available, then rapidly scrambles the mobility spectrum even at relatively small sizes, similarly as in the two-parameter space n, z spanned by naturally charged electrosprays. A simple solution to this problem is to choose the charge-reducing agent to be either A+ or B. This selective charge-reduction process requires a special counter-ion source, being accordingly complex, and limiting the possibilities to freely vary the anion or the cation.

Figure 1. Scheme of the bipolar electrospray chamber with front and back windows; two gray connectors (PEEK) oriented at 90° bringing in one positive and one negative capillary emitter, and gas inlet (right) and outlet (left) connections. A stainless steel grid 30% transparent is symmetrically placed between the capillaries to decouple them electrostatically. An additional image is included in the SI.

Figure 1. Scheme of the bipolar electrospray chamber with front and back windows; two gray connectors (PEEK) oriented at 90° bringing in one positive and one negative capillary emitter, and gas inlet (right) and outlet (left) connections. A stainless steel grid 30% transparent is symmetrically placed between the capillaries to decouple them electrostatically. An additional image is included in the SI.

Figure 2. Bipolar spectra obtained when spraying (a) MC5I+FAP and (b) C7+Br, both dominated by a resolvable series of singly charged clusters A+nBn±1. The original spectrum format (a) is compacted in (b) by using absolute values for the current.

Figure 2. Bipolar spectra obtained when spraying (a) MC5I+FAP− and (b) C7+Br−, both dominated by a resolvable series of singly charged clusters A+nB−n±1. The original spectrum format (a) is compacted in (b) by using absolute values for the current.

Figure 3. High-resolution bipolar mobility spectrum for MC5I+FAP, exhibiting 43 resolvable peaks in each polarity.

Figure 3. High-resolution bipolar mobility spectrum for MC5I+FAP−, exhibiting 43 resolvable peaks in each polarity.

Figure 4. Peak width for the C7+Br negative clusters obtained by Gaussian fitting the spectrum of .

Figure 4. Peak width for the C7+Br− negative clusters obtained by Gaussian fitting the spectrum of Figure 2b.

Figure 5. Resolution (1/FWHM) for the Halfmini DMA challenged with MC5I+FAP (a) and C4+TPB clusters (b), demonstrating resolving powers >50 with sufficiently large particles.

Figure 5. Resolution (1/FWHM) for the Halfmini DMA challenged with MC5I+FAP− (a) and C4+TPB− clusters (b), demonstrating resolving powers >50 with sufficiently large particles.

Figure 6. Bipolar spectrum obtained with the Herrmann DMA for EMI+Methide clusters.

Figure 6. Bipolar spectrum obtained with the Herrmann DMA for EMI+Methide− clusters.

Figure 7. Resolution for the Hermann DMA with EMI+Methide, limited by diffusion in the DMA.

Figure 7. Resolution for the Hermann DMA with EMI+Methide−, limited by diffusion in the DMA.

A considerably simpler source of clusters with only one degree of freedom would be the combination of a positive and a negative electrospray of one given salt. Several variants of the bipolar electrospray have been tried, going back at least to the studies of Borra et al. (Citation1999), and others reviewed recently by Tang (Citation2016). One difficulty encountered is in stabilizing two neighboring oppositely charged electrospray sources, due to counter-ions in the gas (Tang and Gomez Citation1995; Lenggoro et al. Citation2002; Fernandez de la Mora Citation2015). Tang (Citation2016) has short-circuited this difficulty by isolating electrostatically both emitters, each with its own extractor electrode. Since Tang's purpose was to obtain primarily neutral particles, he has introduced an electric field in the neutralization region to drive the two charged clouds against each other. We have used a single grounded screen to isolate electrostatically one electrospray region from a neutralization region containing bipolar ions produced by a radioactive source (Fernandez de la Mora Citation2015). In this way, the ES needle may be brought very close to the screen without any disruption from the neutralizing ions, as the strong electric field prevailing in the vicinity of the spraying capillary does not penetrate through the screen. Our study explored an alternative scenario where the screen was replaced by a small orifice perforated in a separating plate. Two problems arose: first, the spraying tip had to be brought fairly far from the orifice, resulting in substantial spray dilution. Otherwise, ES field leakage through the orifice and penetration of ions from the neutralization chamber destabilized the spray. The effectiveness of the screen approach suggests that the sprayed ions are efficiently propelled through the screen into the neutralization region by their own space charge, enabling a charge-reduction region completely free from electric fields. Space charge self-propulsion then appears to offer relatively large ion lifetimes, and is preferable to Tang's (2016) configuration (an electric field in the neutralization chamber), particularly when the objective is to produce primarily singly charged ions rather than neutrals. TSI's commercial Advanced Aerosol Neutralizer 3088 isolates the spray region from a second neutralization chamber (containing bipolar ions produced by a soft X-ray source) with a metallic plate having a small orifice. This design opposes penetration of the ions into the ES chamber by passing a high-speed gas through the orifice from the ES chamber into the neutralization chamber, permitting on the one hand bringing the ES needle relatively close to the orifice, while also bringing the charged drops very quickly into the charge-reduction region, even before they have a chance to dry. Achieving a high jet speed, however, requires pressurizing the source chamber, which introduces some practical difficulties. In our own design with a small separation plate, we found that the higher resulting jet speed limits the neutralization power relative to that obtained with lower gas speed in the screen approach (Fernandez de la Mora Citation2015). However, this limitation may be overcome with a neutralization chamber large enough for the jet speed to decay.

Given these various precedents, it seemed to us that the proven strategy of isolating with a metallic screen one spray region from another bipolar ion region should be equally effective in isolating two oppositely polarized sprays. Furthermore, space charge should in this case drive even more effectively each of the charged clouds from its original source, through the screen, into the oppositely charged cloud. Once two nearby oppositely polarized sprays are stabilized (with a screen or otherwise) an obvious advantage of this charge-reduction scheme is the great flexibility with which the anion and the cation may be controlled. For instance, by spraying the same salt in the two opposed polarities, only one anion and one cation would be combined (rather than three or more different species, as in conventional gas phase charge-reduction). Then, if the residence time in the charge reduction chamber were controlled to yield mainly singly charged particles, the simple one-parameter series of clusters (BnA+n±1)±1 formed would extend to much larger n values than the currently accessible size range n = 0, 1. The viability of this approach will be demonstrated here.

2. Experimental

2.1. Sprayed solutions

The salts chosen are listed in the right column of . They combine the ions listed in the left column. Corresponding formulae and abbreviations are also given in the table. The two salts including the FAP anion (Ignat'ev et al. Citation2005) and P6,6,6,14+Im originated from Merck, but were obtained indirectly from generous colleagues. C7+Br and C12+Br were from Fluka, EMI+Methide from Covalent and C4+TPB from Sigma.

Table 1. Ions and salts tested.

The positive and negative sprays used exactly the same solution. C4+TPB was dissolved in acetonitrile due to its poor solubility in ethanol. All other salts were dissolved in ethanol. Those that are liquid at room temperature (P6,6,6,14+Im, P6,6,6,14+FAP, MC5I+FAP, EMI+Methide) were at 10−3 volumetric concentrations, and the solid salts at ∼ 2 mM.

2.2. Bipolar ES chamber

sketches the chamber, commercialized by SEADM. It is structurally similar to other ES chambers previously used with a single capillary source, based on a cylindrical body with two windows at its bases, with gas inlet and outlet ports on the sides (see Figure 1 in Fernandez de la Mora Citation2015). The new chamber combines these prior elements with two rather than one liquid chromatography adaptor, enabling introducing two silica capillaries. The capillaries are coplanar with the inlet-outlet axis of the chamber, and are oriented symmetrically at ∼45° with respect to that axis. As the capillaries are pulled from outside the chamber along their axes, the separation between their tips increases, decreasing their mutual electrostatic interaction. The capillaries are made out of silica with inner diameter of either of 75 or 170 μm, and outer diameters of 360 μm (Polymicro Technologies). Their emitting tips are manually pulled under a flame down to a closed cone, cut at the desired capillary tip opening size (30–40 μm), and finally sharpened to a conical tip by rotating them by hand about their axis, while the tip touches lightly a rotating polished flat alumina surface. A stainless steel screen 30% transparent (to light) placed at the plane of symmetry between the two capillaries was used in most experiments. The screen was supported on the gas inlet port. It could be displaced up and down, but its lower end was in most cases slightly downstream the point where the two capillaries intersect. Therefore, even when the two emitters were very close to each other, the two Taylor cones were out of direct sight. Under such conditions, the two sprays were easily stabilized at all positions upstream of the point of intersection of the capillaries. We also examined briefly a scenario without screen, with the capillaries placed as far as possible from each other while being still visible through the window. The sprays were also stable in this configuration, each Taylor cone being barely influenced by changes in the voltage of the other. Considerably more detail on the operation of this source is given in the online supplemental information (SI).

2.3. DMA

Two different cylindrical DMAs of relatively high resolving power at particle diameters of 1 nm were used. One, dubbed the Halfmini, has inner and outer electrode radii R1 = 4 mm; R2 = 7 mm, and an axial distance L = 20 mm between the inlet and outlet slits (Fernandez de la Mora and Kozlowski Citation2013). The other, referred to as a Herrmann DMA is a considerably lengthened (L = 10 cm) modification of a DMA having R1 = 4 mm, R2 = 9 mm, and a wide laminarization trumpet. Shorter versions of the Hermann DMA have been widely used in scientific studies, though without a detailed characterization of its performance. This gap has been recently filled by Kangasluoma et al. (Citation2016a,b), who, however, report a resolving power much inferior to what had been previously shown in the literature (Ku et al. Citation2004; Ku and Fernandez de la Mora Citation2004, Citation2009; Saucy et al. Citation2004; Gamero-Castaño and Fernandez de la Mora Citation2000a,b,c). Most of the exploratory work on the cluster source was done with the Halfmini DMA. The Herrmann DMA was used only in a few experiments (described at the end of the article) meant to characterize its resolving power.

A flow rate qi of approximately 2 L/min (controlled by a flowmeter) of filtered and dried air passes through the bipolar electrospray chamber, entraining the electrosprayed clusters into the polydisperse aerosol inlet to the DMAs. The DMA-selected monodisperse aerosol is captured by an unusually fast (<100 ms response time) and sensitive (0.12 fA noise) Faraday cup electrometer commercialized by SEADM (Fernandez de la Mora et al. Citation2017), which measures the current I carried by the sampled gas. Its flow rate qo is not controlled, but is automatically equal to qi because the sheath gas flow circuit operates in closed loop, with a refrigeration system and a HEPA filter following the blower. The sheath gas flow rate Q was controlled by fixing the speed of rotation of the blower, but was not directly measured. Instead, Q is determined according to relation (Equation1) (Knutson and Whitby Citation1975) from the voltage VC7+ at the center of the most mobile peak appearing in a I(VDMA) spectrum when electrospraying C7+Br:[1]

Z is the electrical mobility under ambient conditions of the C7+ ion used as a mobility standard (ZC7+ = 0.971 cm2/V/s, according to Ude and Fernandez de la Mora Citation2005).

Raw mobility spectra were obtained by fixing all flow rates and the setting of the electrospray source, sweeping the voltage VDMA of the inner electrode while grounding the outer electrode, and recording (in a computer, with the help of a data acquisition card) the aerosol current collected as a function of VDMA. The high voltage power supply used (Applied Kilovolts, ±10 kV, 30 ms response time from +10 to −10 kV) is bipolar and can go continuously through zero (), so bipolar spectra were taken as single data files. In , VDMA is negative for the positive ions and positive for the negative ions. The electrometer is an inverting amplifier giving a negative signal for positive ions. In order to save figure space, the negative currents will be inverted in subsequent figures (), as if the detector were measuring ion concentration.

3. Results

3.1. Dominant production of singly charged clusters

shows the full raw spectrum for MC5I+FAP. Both the negative and positive domains resolve over 40 major peaks associated with singly charged ions. This MC5I+FAP spectrum is taken at limited resolution (Q ∼ 605 L/min) in order to extend the size range and exhibit the transition from individually resolved peaks to just their envelope. Each of the broad modulations seen in the envelope is associated with the original charge state of the clusters prior to charge reduction (Gamero-Castaño and Fernandez de la Mora Citation2000a). is for C7+Br, and is taken at higher resolving power. The negative spectrum resolves 36 major peaks associated with singly charged ions, including a weak peak (∼220 V) of singularly high mobility associated with Br. In addition to the dominant series of ions, much weaker peaks are resolvable in the space between major peaks. Only the monomer A+ and the dimer (A+B)A+ are isolated in conventional unipolar electrospray spectra of alcohol solutions of C7+Br. The trimer (A+B)2A+ typically emerges over a substantial background of multiply charged species, and the tetramer (A+B)3A+ is almost invariably undetectable, overwhelmed by the background of multiply charged particles. The radical shift observed here from the usual undetectable singly charged ions beyond the trimer, into their complete dominance of the spectrum, shows that the charge-reduction process is successful in removing all but the lowest charge states.

The positive spectrum for C7+Br is similarly dominated by singly charged species, although it does not in this case show as many clearly defined peaks as the negative series. The number of peaks that may be distinguished depends considerably on the circumstances of the measurement, especially on the stability of both sprays. This is inevitable given that the largest clusters have mobilities very close to those of their neighbors, so the ripple associated with each individual peak rises only a few percent above the average signal. At high enough n in the A+nBn±1 series, this ripple is no longer observable because of the limited resolution of the Halfmini DMA used. When the neighboring peaks are still resolvable but a slight unsteadiness of the spray current becomes comparable in magnitude to the ripple between peaks, the peaks also become unresolvable, as in the positive spectrum in past n = 27. While a high stability of both sprays is absolutely essential to see the large number of cluster reported in , numerous peaks can nonetheless be observed under less exacting conditions.

A similar pattern of dominantly singly charged clusters has been observed with all the salts tested (, last column). The salt for which the largest number of cluster peaks has been resolved is MC5I+FAP (nmax = 43). The corresponding bipolar spectra are shown in at higher DMA resolving power than in . C16+Br was also tested but did not produce an extended clean cluster series (Figure S2, available in the SI), probably because its minimal solubility in ethanol.

A surprising finding was that some salts produce wider ion mobility peaks than others. This was evident for C12+Br and for the two salts including the Trihexyltetradecylphosphonium cation. This peculiar broadening cannot be due to the DMA, as it happened at exactly the same DMA setting at which other salts gave narrower peaks. Perhaps some clusters have variable shapes and are accordingly not ideal mobility standard. This structural floppiness is well documented in protein ions. It would not be so surprising in the case of solid salts such as C12+Br having very long alkyl branches, which could interfere with the formation of the structure having minimal energy. In the case of the phosphonium ionic liquids, however, assuming that the clusters are also liquid-like, one would expect the cluster to sample all possible shapes millions of times during the DMA measurement, resulting in a unique mobility. Another possible cause of broadening is the presence of contaminants in the salt. This has been documented in mobility mass-spectrometry measurements with large tetra-alkylammonium salts (May et al. Citation2014). It is also evident in some of our measurements. For instance, one sample tried of C4+TPB (Figure S3 in the SI) gives very sharp peaks, many of which are from impurities. The clusters from the dominant series may be resolved from the impurities (up to a point), revealing FWHM as small as 0.02.

3.2. Size range spanned by mobility-resolvable bipolar standards

Various spectra have been so far shown in terms of the DMA voltage, the key quantity governing diffusion broadening and hence limiting resolving power. In order to convert voltage to particle mobility and infer the mobility diameter we have run back-to-back spectra with C7+Br, and the two FAP salts, from which we measure the voltage for the most mobile ion and infer the mobility/voltage calibration data collected in . The flow rate of sheath gas given by Equation (Equation1) was 830 lit/min. It is noteworthy that, in spite of their large masses, the FAP and Im anions are rather compact, with mobilities not far from the polarization limit (∼2 cm2/V/s).

Table 2. Relation between classification voltage and mobility Z for four of the ions used.

The tabulated values have not been obtained precisely enough to be interchangeable with the common C7+ standard, but rather to determine approximately the sizes of the large particles for which pure clusters are resolvable by mobility alone. For example, in , the largest cluster resolved appears at about 4.9 kV, with 1/Z = 1.03*4.9/0.6337 Vs cm−2 = 7.96 Vs cm−2. The corresponding free-molecule mobility diameter dZ = (k/Z)1/2 is 4.03 nm when taking k = 2.04 cm2/V/s/nm2.

3.3. Use of the bipolar ES source for DMA resolution measurements

The development of nanoDMAs with high resolving powers would not have been possible without the availability of molecular standards. However, as we have attempted to extend the range of these instruments beyond their original domain (<10 nm) to 30 nm and beyond, a number of causes conspire to either limit DMA resolving power or preclude its measurement. Longer DMAs classify small ion standards at modest voltages, at which diffusion broadening results in FWHM typically >0.03. Although in an ideal DMA at fixed flow conditions one would expect that peak width would continue decreasing as particle size increases, the lack of larger size standards has precluded confirming this point. Furthermore, as the DMA length increases to accommodate larger particles, it is harder to achieve good electrode centering, particularly in the aerodynamically favorable Reischl (Citation1991) geometry, where the central electrode is supported only on the downstream end. While long DMAs may support their inner electrode on both ends and also have aerodynamically favorable sheath gas inlets, the likelihood of turbulent transition also increases with increased instrument length. As a result, in our recent work with proteins and viral particles, we have been unable to ascertain if the finite peak widths found were intrinsic to these particles, or were rather due to nonideal DMA response. In fact, the maximum resolving power so far reported for the long (2 cm) Halfmini DMA is in the range 30–40, and is limited by the previously available size standards. As seen in , resolvable peaks are now generated by the bipolar source at mobilities about 10 fold those previously available, enabling in principle a reduction by 101/2 ( = 3.16) the diffusion-limited peak width.

3.3.1. HalfMini DMA

We have determined peak widths by fitting the spectra previously presented to a linear combination of Gaussians. plots FWHM for negative C7+Br clusters in the spectrum of . Their width decreases initially with DMA voltage, but exhibits an unnatural transition above 2 kV. This is surely an artifact following from incomplete separation of the peaks above that voltage. Indeed, the Gaussian inversion has been performed assuming the complete absence of a broad background of impurities under the partially resolved peaks, which artificially increases their height and the inferred width. In spite of this limitation, the Figure demonstrates a resolving power greater than 50, unprecedented for a DMA whose length over gap ratio L/(R2-R1) = 6.67 exceeds considerably the ideal ratio of 1.

The spectra of C4+TPB (Figure S4) have been analyzed in more carefully than those of C7+Br. In this case, we have included resolvable impurity ions in the Gaussian fitting, so as to remove their interference. As an additional precaution, the widths of secondary peaks have not been used to determine resolving power. This more thorough analysis confirms that the resolution exceeds 50. plots the resolution versus VDMA1/2 because the ideal diffusion broadened resolution is a straight line through the origin (Tammet Citation1970; Stolzenburg Citation1988; Rosell et al. Citation1996), facilitating a visual extrapolation of the data. No comparison with theory is provided in that figure because the performance always falls below the pure diffusion limit. Since the level of peak interference and the noise in inferred resolution become substantial as the resolving power reaches 50, it would be dangerous to extrapolate this trend. The cleaner mobility spectra obtained with MC5I+FAP () have resulted in the peak width measurements of , demonstrating for the first time that this DMA has regimes reaching a resolution of 52–54. Although peak interference precludes reliable resolution inferences beyond this value, the slope of the curve shows no sign of saturation. The resolving power of this DMA is clearly better than 52, but the bipolar clusters available are not capable of measuring it beyond this value. While changes in the flow rates over the very large range of values open to this instrument may lead to drastic variations over this performance, this limited result under even one condition provides complete assurance that the concentricity of the instrument is not a limiting feature.

3.3.2. Herrmann DMA

This is the first prototype ever built of a Herrmann DMA with an axial length as large as 10 cm. It was modified from a shorter Herrmann DMA to study viral particles over the size range up to 60 nm, a development still in progress. Given the wide size range of this DMA it had previously not been possible with conventional cluster standards to establish if its resolving power exceeded even 18. This ambiguity has frustrated our goal of separating virus particles with a resolution exceeding 30. Given the limited available information on the intrinsic mobility peak width for viruses following electrospray charge reduction, it has remained unclear if one needs to focus on improving the DMA, the charge reduction process, the solution cleaning protocol, etc.

The new bipolar source has greatly clarified these doubts. shows a bipolar spectrum for an electrospray of the EMI+Methide ionic liquid, including 25 resolved ions in the positive spectrum. shows the corresponding resolving power, reaching up to 33 with no indication of saturation with increasing particle size.

4. Conclusions

Electrostatic isolation with a metallic grid of two oppositely polarized electrosprays enables their stabilization even when the emitting tips are very close to each other. Interaction of the two oppositely charged sprays then leads to efficient charge reduction, resulting in a bipolar aerosol dominated by singly charged clusters.

This behavior is as expected based on prior experience with a charge-reduction system where a metallic grid separated electrostatically the spray region from a bipolar ion region. One substantial advantage of the bipolar electrospray is that, when the same salt solution is sprayed from both capillaries, the clusters formed are members of the simple series of pure species (An+Bn±1)±1, defined by the single parameter n. Many members of this series can be resolved by mobility alone, offering a simple means to provide an aerosol of nanoparticles with essentially pure composition. By classification with a Halfmini DMA the cluster size may be changed from a single molecular ion up to 4 nm. Use of a DMA with a superior resolving power would extend this maximum resolvable cluster size. We have also attempted to increase the maximal resolvable size by operating with salts composed of unusually large anions and cations. The few tested here involving C16+, C12+, and the P6,6,6,14+ phosphonium cation have not been completely successful, perhaps due to the lack of either a rigid structure or sufficient sample purity. We trust this strategy will be fruitful if more systematically pursued.

The MC5I+FAP salt is the one so far having yielded the narrowest mobility distributions, with over 43 positive clusters resolvable reaching particle sizes up to 4 nm.

These standards have been exploited here to characterize two nanoDMAs designed to go beyond the 10 nm size range previously characterizable with available 1 nm size standards. The 2 cm long Halfmini DMA has been seen to achieve resolving powers beyond 50, considerably larger than previously considered possible. The 10 cm Herrmann DMA designed for viral particle analysis has a resolving power of at least 33, which continues increasing monotonically with particle size.

Conflict of interest statement

Juan Fernandez de la Mora declares a personal interest in the company SEADM commercializing the DMA, the electrometer and the bipolar electrospray source used in this work.

Supplementary materials

The supplementary information associated with the present article provides bipolar mobility spectra for a wide group of salts listed in Table 1, as well as details on the following subjects: overall response of the bipolar source; relevance of fast response detector. The bipolar electrospray as a source of pure neutral clusters; relative transmission of the Herrmann and the Halfmini DMAs.

Supplemental material

UAST_1302070_Supplemental_File.zip

Download Zip (780.7 KB)

Acknowledgments

We thank Mr. Mario Amo-Gonzalez of SEADM for his valuable inputs in the design of the bipolar source chamber. We are indebted to Prof. Michel Attoui (Paris) and Luis J. Perez-Lorenzo (Yale) for numerous contributions to the experimental setup used to characterize the bipolar chamber.

References

  • Attoui, A., Paragano, M., Cuevas, J., and Fernandez de la Mora, J. (2013). Tandem DMA Generation of Strictly Monomobile 1–3.5 nm Particle Standards. Aerosol Sci. Techno., 47(5):499–511.
  • Borra, J. P., Camelot, D., Chou, K.-L., Kooyman, P. J., Marijnissen, J. C. M., and Scarlett, B. (1999). Bipolar Coagulation For Powder Production: Micro-Mixing Inside Droplets. J. Aerosol Sci., 30(7):945–958.
  • Fenn, J. B., Mann, M., Meng, C. K., Wong, S. F., and Whitehouse, C. M. (1989). Electrospray ionization for mass spectrometry of large biomolecules. Science, 246:64–71
  • Fernandez de la Mora, J., and Kozlowski, J. (2013). Hand-Held Differential Mobility Analyzers of High Resolution for 1-30 nm Particles: Design and Fabrication Considerations. J. Aerosol Sci., (57):45–53.
  • Fernandez de la Mora, J. (2015). High-Resolution Mobility Analysis of Charge-Reduced Electrosprayed Protein Ions. Anal. Chem., 87:3729–3735.
  • Fernandez de la Mora, J., Perez-Lorenzo, L. J., Arranz, G., Amo-Gonzalez, M., and Burtscher, H. (2017). Fast high-resolution nanoDMA measurements with a 25 ms response time electrometer. Aerosol Sci. Technol., in press. doi: 10.1080/02786826.2017.1296928
  • Gamero-Castaño, M., and Fernandez de la Mora, J. (2000a). Modulations in the Abundance of Salt Clusters in Electrosprays. Anal. Chem., 72:1426–1429.
  • Gamero-Castaño, M., and Fernandez de la Mora, J. (2000b). Mechanisms of Electrospray Ionization of Singly and Multiply Charged Salt Clusters. Anal. Chimica Acta., 406:67–91.
  • Gamero-Castaño, M., and Fernandez de la Mora, J. (2000c). Kinetics of Small Ion Evaporation from the Charge and Size Distributions of Multiply Charged Electrospray Clusters. J. Mass Spectrometry, 35:790–803.
  • Gamero-Castaño, M., and Fernandez de la Mora, J. (2000d). A Condensation Nucleus Counter (CNC) Sensitive to Singly Charged Subnanometer Particles. J. Aerosol Sci., 31:757–772.
  • Kangasluoma, J. (2015). Generation, Characterization and Sizing of Sub-3 nm Nanoparticles and Molecular Clusters, Ph.D. Thesis, Department of Physics, University of Helsinki.
  • Ignat'ev, N. V., Welz-Biermann, U., Kucheryna, A., and Bissky, G. (2005). New Ionic Liquids with tris(perfluoroalkyl)trifluorophosphate (FAP) Anions. J. Fluo. Chem., 126(8):1150–1159.
  • Jiang, J. K., Zhao, J., Chen, M. D., Eisele, F. L., Scheckman, J., Williams, B. J., and McMurry, P. H. (2011). First Measurements of Neutral Atmospheric Clusterand 1–2 nm Particle Number Size Distributions during Nucleation Events. Aerosol Sci. Technol., 45(4):ii–v.
  • Kane, D., Daly, G. M., and El-Shall, S. (1995). Condensation of Supersaturated Vapors on Benzene Ions Generated by Resonant 2-Photon Ionization, a New Technique for Ion Nucleation. J. Phys. Chem., 99:7867–7870.
  • Kane, D., and El-Shall, S. (1996). Ion Nucleation as a Detector: Application of REMPI to Generate Selected Ions in Supersaturated Vapors. Chem. Phys. Lett., 259:482–487.
  • Kangasluoma, J., Attoui, M., Korhonen, F., Ahonen, L., Siivola, E., and Petäjä, T. (2016a). Characterization of a Herrmann-Type High-Resolution Differential Mobility Analyzer. Aerosol Sci. Tech., 50(3):222–229.
  • Kangasluoma, J., Samodurov, A., Attoui, M., Franchin, A., Junninen, H., Korhonen, F., Kurtén, T., Vehkamäki, H., Sipilä, M., Lehtipalo, K., Worsnop, D. R., Petäjä, T., and Kulmala, M. (2016b). Heterogeneous Nucleation onto Ions and Neutralized Ions: Insights into Sign-Preference. J. Phys. Chem. C., 120:7444–7450.
  • Kaufman, S. L., Skogen, J. W., Dorman, F. D., Zarrin, F., and Lewis, K. C. (1996). Anal. Chem., 68:1895–1904.
  • Kaufman, S. L., and Dorman, F. D. (2008). Sucrose Clusters Exhibiting a Magic Number in Dilute Aqueous Solutions. Langmuir., 24(18):9979–9982.
  • Knutson, E. O., and Whitby, K. T. (1975). Aerosol Classification by Electric Mobility: Apparatus Theory and Applications. J. Aerosol Sci., 6:443–451.
  • Ku, B. K., Fernandez de la Mora, J., Saucy, D. A, and Alexander, J. N. (2004). Mass Distribution Measurement of Water-Insoluble Polymers by Charge-Reduced Electrospray Mobility Analysis. Anal. Chem., 76:814–822.
  • Ku, B. K., and Fernandez de la Mora, J. (2004). Cluster Ion Formation in Electrosprays of Acetonitrile Seeded with Ionic Liquids. J. Phys. Chem. B., 108:14915–14923.
  • Ku, B. K., and Fernandez de la Mora, J. (2009). Relation between Electrical Mobility, Mass, and Size for Nanodrops 1–6.5 nm in Diameter in Air. Aerosol Sci. & Technol., 43(3):241–249.
  • Lenggoro, I.W., Xia, B., Okuyama, K., and Fernandez de la Mora, J. (2002). Sizing of Colloidal Nanoparticles by Electrospray and Differential Mobility Analyzer Methods. Langmuir., 18(12):4584–4591.
  • May, J. C., Goodwin, C. R., Lareau, N. M., Leaptrot, K. L., Morris, C. B., Kurulugama, R. T., Mordehai, A., Klein, C., Barry, W., Darland, E., Overney, G., Imatani, K., Stafford, G. C., Fjeldsted, J. C., and McLean, J. A. (2014). Conformational Ordering of Biomolecules in the Gas Phase: Nitrogen Collision Cross Sections Measured on a Prototype High Resolution Drift Tube Ion Mobility-Mass Spectrometer. Anal. Chem., 86:2107–2116.
  • Reid, G. E., Wells, J. M., Badman, E. R., and McLuckey, S. A. (2003). Int. J. Mass Spectrom., 222:243–258.
  • Reischl, G. P. (1991). Measurement of Ambient Aerosols by the DifferentialMobility Analyzer Method: Concepts and Realization criteria for the SizeRange between 2 and 500 nm. Aerosol Sci. Technol., 14:5–24.
  • Rosell, J., Loscertales, I. G., Bingham, D., Fernández de la Mora, J. (1996). Sizing Nanoparticles and Ions with a Short Differential Mobility Analyzer. J. Aerosol Sci., 27:695–719.
  • Saucy, D., Ude, S., Lenggoro, W., and Fernandez de la Mora, J. (2004). Mass Analysis of Water-Soluble Polymers by Mobility Measurement of Charge-Reduced Electrosprays. Anal. Chem., 76:1045–1053.
  • Steiner, G., Franchin, A., Kangasluoma, J., Kerminen, V.-M., Kulmala, M., and Petäjä, T. (2017). Production of neutral molecular clusters by (controlled) neutralization of mobility standards. Aerosol Sci. & Tech., under review.
  • Stolzenburg, M. R. (1988). An Ultrafine Aerosol Size Distribution Measuring System. Ph.D. Thesis, University of Minnesota, Minneapolis, MN].
  • Tammet, H.F. (1970). The Aspiration Method for the Determination of Atmospheric-ion Spectra. Israel Program for Scientific Translations. Jerusalem. The original work in Russian was published in 1967.
  • Tang, J. (2016). Application of Electrospraysfor Nanomaterial Deposition and Synthesis for Energy Conversion and Storage Devices, Ph.D. Thesis, Yale University.
  • Tang, K., and Gomez, A. (1995). Generation of Monodisperse Water Droplets from Electrosprays in a Corona-Assisted Cone-Jet Mode. J. Colloid Interface Sci., 175(2):326–332.
  • Ude, S., and Fernández de la Mora, J. (2005). Molecular Monodisperse Mobility and Mass Standards from Electrosprays of Tetra-Alkyl Ammonium Halides. J. Aerosol Sci., 36(10):1224–1237.
  • Winkler, P. M., Vrtala, A., Steiner, G., Wimmer, D., Vehkamäki, H., Lehtinen, K. E. J., Reischl, G. P., Kulmala, M., and Wagner, P. E. (2012). Quantitative Characterization of Critical Nanoclusters Nucleated on Large Single Molecules. Phys. Rev. Lett., 108:085701.
  • You, R., Li, M., Guha, S., Mulholland, G. W., and Zachariah, M. R. (2014). Bionanoparticles as Candidate Reference Materials for Mobility Analysis of Nanoparticles. Anal. Chem., 86:6836–6842.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.