High-Resolution Ct of the Lung in Asbestos-Exposed Subjects

Comparison Of Low-Dose And High-Dose Hrct

, , &
Pages 473-477 | Accepted 21 Dec 1993, Published online: 04 Jan 2010
 

Abstract

The lowest possible mAs settings for high-resolution CT (HRCT) were studied on 45 individuals with suspected asbestos-related lung disease. All patients were investigated with 5 to 6 high-dose HRCT images (120 kVp/160 mA/2 s) at 3-cm intervals. At a selected level 4 additional low-dose images were obtained on each patient with lower mAs settings (100 mA/2 s, 80 mA/2 s, 60 mA/2 s, 30 mA/ 2 s). Thirty-seven subjects out of 45 had HRCT lesions compatible with asbestosis. HRCT images obtained with as low as 60 mA/2 s settings clearly showed pleural tractions and thickenings, parenchymal bands, honeycombing and subpleural curvilinear shadows, whereas in the evaluation of subpleural short lines and ground glass findings 80 mA/2 s were required. The lowest setting, 30 mA/2 s, was sufficient only in detecting and evaluating pleural tractions and thickenings. We conclude that 160 mAs yield good quality HRCT images, with substantial decrease of radiation dose, for the evaluation of asbestos-related lesions.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.