Reliability of Radiological Measurements of the Distraction Gap During Leg Lengthening

, , &
Pages 162-165 | Accepted 27 Apr 1995, Published online: 04 Jan 2010
 

Abstract

Purpose: The aim of this study was to determine the accuracy of length measurements on plain radiographs during leg lengthening and to evaluate a reliable method of measuring the amount of lengthening suitable for clinical praxis.

Material and Methods: In an experimental study, a plastic femur was lengthened 40 mm and assessed radiographically in different positions, and the orthoradiographically calculated total length of the femur was compared with the true value. In a clinical study, 96 radiographs encompassing 14 femoral and 2 tibial lengthenings were assessed on a digitized table with a computer program (PROFILE) with regard to the amount of lengthening and the degree of magnification.

Results: The radiographic magnification in the experimental study, depending on position of the femur, was 17% (5–25%), and in the clinical material it averaged 16% (0–36%).

Conclusion: Radiographs obtained with a radiopaque ruler placed at the same level as the lengthened bone allowed reliable measurements of the distraction gap.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.