Phantom Gastric Mucosa for Evaluating Computed Radiography in Double-Contrast Upper Gastrointestinal Examinations

, &
Pages 204-207 | Accepted 06 Jun 1995, Published online: 04 Jan 2010
 

Abstract

Purpose: To test the clinical usefulness of computed radiography (CR) with a storage phosphor plate in upper gastrointestinal radiographic examinations, a newly devised phantom gastric mucosa was used.

Material and Methods: Simulated small elevated and depressed lesions were created on a phantom gastric mucosa made from a styrofoam “plate”. Twenty-four sets of each CR and screen-film radiographs (SR) were obtained using phototimed exposures. Receiver operating characteristic (ROC) study and visual ranking using these images were performed.

Results: There was no significant difference between the ROC curves of CR and SR. By visual ranking, CR was equal to or better than SR in most cases. In no case was SR definitely superior to CR.

Conclusion: CR can be safely applied in upper gastrointestinal roentgenologic examinations.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.