ACCURACY OF ULTRASONOGRAPHY AND MAGNETIC RESONANCE IMAGING IN DETECTING FAILURE OF BREAST IMPLANTS FILLED WITH SILICONE GEL

Pages 415-418 | Published online: 08 Jul 2009
 

Abstract

We prospectively studied the accuracy of magnetic resonance imaging (MRI) and ultrasonography (US) for preoperative detection of rupture in 35 single-lumen implants filled with silicone gel in 18 patients. The positive predictive value of US for rupture of an implant was 70% and the negative predictive value 64%. Sensitivity and specificity were 44% and 87%, respectively. Accuracy, defined as the total true positive and true negative values divided by the total number of implants studied was 66%. The positive predictive value of MRI was 100% and the negative predictive value 90%. The corresponding sensitivity and specificity were 88% and 100% and the accuracy 94%. MRI offers significantly better diagnostic sensitivity (p = 0.02) and accuracy (p = 0.004), and should be regarded as the ?gold standard? in the evaluation of rupture of breast implants filled with silicone gel. When MRI is not readily available, US is an acceptable alternative.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.