Abstract
Trace amounts (ppb or less) of phosphorus containing compounds present in aqueous samples are adsorbed on XAD-4 and subsequently eluted by means of ethyl acetate. The solvent and the eluted compounds are evaporated and swept over a Tenax-GC tube. This gas stripping method traps the phosphorus containing compounds together with only a small amount of the solvent whereas the water entrapped in the XAD step is removed simultaneously. The compounds are desorbed from the Tenax-GC tube and injected into the gas chromatograph using the combination of thermal desorption, cold trapping and flash heating. The subsequent analysis is carried out on a capillary column and the compounds are detected by means of a flame photometric detector. The various steps of the analytical procedure are discussed, including the recoveries of the different compounds studied and some instrumental aspects.