Neonatal life support during magnetic resonance imaging

, , &
Pages 71-74 | Published online: 04 Aug 2009
 

Abstract

Magnetic resonance techniques are required frequently for the assessment of the brain of ill neonates. In the present study, the effects of a 1.5 T MR scanner on devices for life support were assessed. A ventilator (Dra¨ger Babylog 2000) was tested in the 1.5 T magnet, using a neonatal ventilation tester and 1.5-5 m tubes. In a special MR incubator, temperature and humidity were measured at 1-min intervals. Infusion was tested with the pump outside the magnet room: infusion rates and time to alarm were tested with 7-m tubes. The ventilator performed normally at a magnetic field line of 2 mT, although the alarms failed. The incubator created a temperature of 35.9°C and humidity of 40.7%, which was acceptable for examinations of 45 min. The alarm limits of the infusion pump placed outside the magnet at 7 m were within company limits. The study indicates that magnetic resonance examinations can be performed safely in ill preterm neonates who require life-support devices.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.