Effect of electrode impedance in improved buffer amplifier for bioelectric recordings

, , , &
Pages 217-222 | Published online: 09 Jul 2009
 

Abstract

We analysed the effects of electrode impedance on the transfer response of a one-stage improved buffer amplifier. The electrode DC resistance (Rd) modifies the one-stage buffer transfer response. We found a limit electrode resistance (Rd(lim)) which depends on the transfer damping factor (ϵ). If Rd is lower than 86.5 k Ω, the transfer response of the buffer fulfils American Heart Association (AHA) recommendations, but when Rd is greater than Rd(lim) it must be cautiously weighed up because its influence in the transfer response becomes appreciable. The maximum Rd that can be driven by the buffer is 1.2 M Ω. Higher values do not fulfil AHA recommendations. Therefore, electrodes with higher impedance should not be used with this kind of buffer. In contrast, when this buffer is used to build in an instrumentation amplifier (IA) for bipolar recording, the common-mode rejection ratio (CMRR) is sensitive to the electrode type used.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.