World-smallest LVAD with 27 g weight, 21 mm OD and 5 l min−1 flow with 50 mmHg pressure increase

, , , , , & show all
Pages 181-184 | Published online: 09 Jul 2009
 

Abstract

To investigate the feasibility of a long-term left ventricular assist device (LVAD) placed in the aortic valve annulus, an implantable aortic valve pump (21 mm outer diameter, weighing 27 g) was developed. The device consists of a central rotor and a stator. The rotor assembly incorporates driven magnets and an impeller. The stator assembly has a motor coil with an iron core and outflow guide vanes. The device is to be implanted identically to an aortic valve replacement, occupying no additional anatomic space. The pump delivers the blood directly from left ventricle to the aortic root, like a natural ventricle, therefore causing less physiologic disturbance to the natural circulation. Neither connecting conduits nor ‘bypass’ circuits are necessary. The pump is designed to cycle between a peak flow and zero net flow to approximate systole and diastole. Bench testing indicates that the pump can produce a blood flow of 5 l min−1 with 50 mmHg pressure increase at 17 500 rpm. At zero net flow rate, the pump can maintain a diastole aortic pressure against 80 mmHg at the same rotating speed.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.