Low Flow Rate Sharp Cut Impactors for Indoor Air Sampling: Design and Calibration

, , &
Pages 1303-1307 | Received 09 Jan 1987, Accepted 31 Jul 1987, Published online: 08 Mar 2012
 

Abstract

Two single round nozzle impactors have been developed for use in Harvard’s indoor air pollution health study. Both impactors operate at flow rates of 4 L/m and are nearly identical, differing only in their cut sizes of 2.5 μm and 10 μm aerodynamic diameters. Two identical cascaded stages of the same cut size are used to obtain sharp cut-off characteristics. The particles are deposited on impaction plates made of oil impregnated, porous material to reduce particle bounce and are discarded. Only the particles collected on the afterfilter are analyzed. Special care has been taken to collect the particles uniformly on the afterfilter to aid in particle analysis.

The jmpactors were calibrated with a vibrating orifice monodisperse aerosol generator. However, due to the sharp cut of the impactors, doublets and triplets in the calibration aerosols, even in small quantities, gave erroneous calibration curves. Therefore, the number of doublets and triplets in the challenging aerosols were measured and appropriate corrections made to the calibration curves.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.