Abstract
This case describes some of the commonly overlooked device-related issues in patients who have reportedly failed to respond to cardiac resynchronization therapy (CRT). The case demonstrates voltage-dependent right ventricular capture instead of right atrial capture by a subtly malpositioned right atrial lead. CRT therapy failed to improve symptoms of heart failure and the diagnosis of “CRT nonresponder” was made. With a detailed fact-finding approach, the mechanism behind this nonresponse was identified, and the outcome of CRT was significantly improved with rectification of the problems.