Noninvasive Instrument-based Tests for Detecting and Measuring Vitreous Inflammation in Uveitis: A Systematic Review

, MBChBORCID Icon, , , MBChB, , MPH, , MSc, , MD, , PhD & , PhDORCID Icon show all
Pages 137-148 | Received 08 Jun 2020, Accepted 17 Jul 2020, Published online: 06 Oct 2020

ABSTRACT

Purpose

This systematic review aims to identify instrument-based tests for quantifying vitreous inflammation in uveitis, report the test reliability and the level of correlation with clinician grading.

Methods

Studies describing instrument-based tests for detecting vitreous inflammation were identified by searching bibliographic databases and trials registers. Test reliability measures and level of correlation with clinician vitreous haze grading are extracted.

Results

Twelve studies describing ultrasound, optical coherence tomography (OCT), and retinal photography for detecting vitreous inflammation were included: Ultrasound was used for detection of disease features, whereas OCT and retinal photography provided quantifiable measurements. Correlation with clinician grading for OCT was 0.53–0.60 (three studies) and for retinal photography was 0.51 (1 study). Both instruments showed high inter- and intra-observer reliability (>0.70 intraclass correlation and Cohen’s kappa), where reported in four studies.

Conclusion

Retinal photography and OCT are able to detect and measure vitreous inflammation. Both techniques are reliable, automatable, and warrant further evaluation.

Vitreous inflammation, or vitritis, is a clinical manifestation commonly found in posterior-segment involving uveitis. It is the hall-mark of intermediate uveitis, but is also common in panuveitis and may occur in posterior uveitis.Citation1,Citation2 Infiltration of the vitreous body with inflammatory cells and proteinaceous exudates gives a characteristic hazy appearance, reducing the clarity of structures behind it (the optic disc and retinal vessels) during fundoscopy.Citation3 The clinical standard for measuring vitreous haze has been the National Eye Institute vitreous haze (NEI VH) scale since the Standardization of Uveitis Nomenclature (SUN) Workshop in 2005.Citation1 Prior to the SUN workshop, three grading systems existed.Citation2,Citation4,Citation5 The NEI VH scale is a 6-point grading system for estimating the vitreous clarity as seen through indirect ophthalmoscopy and is also referred to as the National Institute for Health (NIH) or Nussenblatt scale.Citation1,Citation2 The clinician’s estimate is compared to a standardized set of photographs and given a score of 0, +0.5, +1, +2, +3, or +4 (). This grading system has been the widely accepted standard for clinical assessment in routine care and for assessing disease outcomes in clinical trials.Citation6Citation8 It has been adopted as part of composite measures of disease outcome for uveitis, alongside other markers of inflammation such as anterior chamber cells/flare, central macular thickness, visual function, and quality of life.Citation9,Citation10 However, there are drawbacks to clinician grading. Firstly, this method is subjective with only moderate interobserver agreement, even when assessed by experienced uveitis specialists.Citation11,Citation12 Secondly, the grading scale is non-continuous and non-linear, with large steps between each grade. Lastly, the system is poorly discriminatory for low levels of vitreous inflammation, where the need for sensitive detection of inflammatory activity to allow early clinical intervention, is greatest.Citation13

Table 1. Standardization of uveitis nomenclature/Nussenblatt photographic grading of vitreous hazea

More recently, measuring vitreous inflammation using instrument-based systems such as imaging devices has been proposed as a solution to some of these challenges. Several instruments, including fundus photography, ultrasound, and optical coherence tomography (OCT) have been used to visualize the vitreous body. These instrument-based methods have the theoretical advantage of being objective and automatable, and the changes detectable by each could be employed as surrogate measures of vitreous inflammation. This systematic review aims to identify all non-invasive, instrument-based tools (hereon referred to as index tests) with the ability to detect and measure vitreous inflammation in uveitis, and report the level of correlation between index tests and clinician grading, as well as the index tests’ reliability.

Methods

This systematic review is reported according to the Preferred Reporting Items for Systematic Reviews and Meta-Analysis (PRISMA) statement.Citation14 The methodology was specified in advance and the protocol registered with PROSPERO (CRD42017084168).Citation15 Our search seeks to identify all index tests for detecting and quantifying vitreous inflammation. Where index tests were compared against a clinician grading system, the level of correlation was extracted. Any evaluation of test reliability, such as intra- and inter-observer reliability was also extracted.

Search strategy

We combined free text terms and index terms reflecting the pathological finding of interest, ‘vitreous haze’ or ‘vitritis’ and the disease context ‘uveitis,’ ‘inflammation,’ ‘blood-retinal barrier,’ and ‘leak’ where possible (search strategy available in Supplementary Materials). Database searches were carried out in MEDLINE, Embase, Cochrane Controlled Register of Trials (CENTRAL), Center for Reviews and Dissemination Database (Health Technology Assessments and the Database of Abstracts and Reviews of Effects), Clinicaltrials.gov, WHO International Clinical Trials Registry Platform (ICTRP portal), British Library’s ZETOC, Conference Proceedings Citation Index (Web of Science), British Library Ethos, ProQuest and OpenGrey. We searched all databases from inception to December 4, 2019, with no date or language restrictions. We manually searched citations of review articles and included studies to identify additional relevant articles.

Study selection

Two reviewers independently assessed study eligibility and resolved disagreements by consensus or by referral to a third reviewer. Studies were eligible if they described one or more index tests for detecting and measuring vitreous inflammation. Studies were not excluded based on the basis of subject age, gender, ethnicity, underlying etiology, or disease activity status. Animal studies and studies involving only healthy participants, single case reports, commentaries, and opinion articles were excluded.

Data extraction

Two reviewers independently extracted data using a pre-specified data extraction sheet and resolved any discrepancies through consensus and referral to a third reviewer when needed. Data extracted included study design, population characteristics and disease phenotype, details of the index and reference tests, and outcomes relating to correlation between the two tests and test reliability. The full list of extracted items can be found in .

Table 2. Study characteristics.

Table 2. (Continued).

Table 2. (Continued).

Risk of Bias Assessment

Relevant features of the Quality Assessment of Diagnostic Accuracy Studies tool (QUADAS-2) were used to assess for bias in the studies. The assessment considered patient selection (if the patients receiving the index and reference tests were representative of uveitis patients and the spectrum of uveitic subtypes), index test (if the index test was interpreted without knowledge of the reference test), reference test (if the reference test was interpreted without knowledge of the index test) and flow and timing (if all patients received both tests within an appropriate time interval – within same day assessment was deemed sufficient to ensure the inflammatory status of the eye had not changed). Not all elements of QUADAS-2 were applicable. For example, “whether the reference standard is likely to correctly quantify the target disease (vitreous inflammation)” would be marked unclear for all studies, due to the known poor reliability of clinician grading. As QUADAS-2 is only applicable for studies comparing an index test to reference test, the assessment was only carried out in studies evaluating correlation between the two tests and not in studies evaluating index test reliability.

Data analysis

For each index test, we tabulated the extracted information and provided a narrative synthesis of methodological characteristics and index tests evaluated. Studies which compared index test measurements with a reference test (such as clinician grading) and reported a correlation coefficient were included in the analysis. In these studies, where confidence intervals for correlation coefficients were not reported, correlation coefficients were normalized using Fisher’s Z transformation for meta-analysis and back transformed and presented on a forest plot for visualization only. All statistical analyses were performed using Stata Statistical Software (Release 15. College Station, TX: StataCorp LP). Meta-analysis was not performed for test correlation or reliability due to heterogeneity between studies.

Results

Results of the Search

The study selection process is summarized in the PRISMA flow diagram ().

Figure 1. PRISMA flow diagram.

Figure 1. PRISMA flow diagram.

The search yielded 7122 unique bibliographic records after removal of duplicates. Of these, 7100 were excluded based on screening of titles and abstracts. The large number of excluded records was due to the unrestrictive nature of the search strategy, which was deliberately wide, to ensure full capture of all potentially relevant technologies. The remaining 22 articles were reviewed in full text and further 10 articles were excluded. The reasons for exclusion were due to not matching the criteria for outcome (n = 6) or target population (n = 4). Twelve articles were included; two studies compared an index test with a clinician grading system, two reported test reliability, and two did both. Six studies described index tests but did not report correlation with clinician grading or index test reliability ().

Participants’ characteristics and study design

The 12 studies included a total of at least 840 participantsCitation16-Citation25 (two studies did not report the number of participantsCitation26,Citation27) and at least 846 eyes (one study did not report the number of eyes.Citation23) The studies were published between 1977 and 2019. Four studies were conducted prospectivelyCitation16,Citation17,Citation24,Citation25 and eight were retrospective.Citation18-,Citation23-,Citation26-,Citation27 Only five studies reported gender, with 29% of participants (n = 149) being male.Citation16,Citation17,Citation22-Citation24 The age of participants in studies ranged from 12 to 75 years.Citation16,Citation17,Citation19-Citation25 Five studies included mixed etiologies, including toxoplasmosis, sarcoidosis, Behcet’s disease, Birdshot chorioretinopathy, pars planitis, Vogt-Koyanagi-Harada disease, multiple sclerosis, tubulointerstitial nephritis and uveitis syndrome, Reiter’s syndrome, multifocal retinitis, serpiginous choroiditis, idiopathic retinal vasculitis, sympathetic ophthalmia, and Dengue retinitis.Citation19Citation22,Citation24 Two studies were narrower in their inclusion criteria with one study including intermediate uveitis onlyCitation17 and the other including patients with multiple sclerosis onlyCitation23. Five studies did not specify the underlying etiology of participants.Citation16,Citation25-Citation27

Clinical reference test

Six out of 12 studies did not compare an index test against a comparator.Citation17,Citation22-Citation26 One study compared ultrasound biomicroscopy against qualitative features on slit-lamp fundoscopy.Citation16 Four studies compared OCTCitation19-Citation21 and retinal photographyCitation27 against the NEI VH scale only, and one study compared retinal photography against both NEI VH scale and a photographic scale called the Miami scale (described in next section).Citation27

Instruments for detecting and quantifying vitreous inflammation

Three types of technologies with the ability to detect and quantify vitreous inflammation were identified from the 12 studies: ultrasound, retinal photography, and OCT.

Three studies employed ultrasound. One study used an A-scan instrument, model 7100A (Kretztechnik, Austria) with a transducer of 6 MHz/S mm25 and two studies used ultrasound biomicroscopy (the UBM 840 (Zeiss-Humphrey, San Leandro, CA, USA) with a 50 MHz probe in one studyCitation16 and the Model P45 (Paradigm Medical Industries, Salt Lake City, Utah, USA) with a 50 MHz probe plus the Cinescan S (Quantel Medical, Clermonth-Ferrand, France) with a 20 MHz immersion open probe in another study.Citation17) The images in all three studies were interpreted manually and qualitatively, by the operator, in real-time.

Three studies used retinal photography. Davis et al. developed a 9-point scale using calibrated Bangerter filters to blur fundus photographs, originally acquired using 30⁰ Zeiss fundus camera model FF4 (Carl Zeiss Meditec Inc, Pleasanton, California, USA) with a Nikon film camera (Nikon Instruments Inc, Melville, New York, USA).Citation26 This 9-point scale is known as the Miami scale and is designed to be a reference for manual clinician grading of fundus photographs. The authors tested the use of this reference scale using film fundus photographs from an imaging archive (unspecified camera and system). Madow et al. used fundus photographs originally acquired as color film slides for the MUST trialCitation12 and digitized them using Nikon Coolscan film scanner (Nikon, Inc, Melville, New York, USA) at 300 dpi and saved as TIFF format.Citation18 Madow et al. used the Miami scale developed by Davis et al. to grade the severity of vitreous haze in these photographs.Citation18 Passaglia et al. applied an automated retinal photography analysis software to grade fundus photographs from a clinical trial library (unspecified source, camera, and system) according to the NEI VH and Miami VH scales.Citation27

Six studies used OCT. Five studies used the Heidelberg SPECTRALIS OCTCitation19,Citation20,Citation22-Citation24 and one used the Cirrus HD-OCT (Carl Zeiss Meditec, Dublin, California, USA).Citation21 Two studies used the same semi-automated image analysis technique (custom OCTOR software),Citation19,Citation22 two used the same fully automated image analysis technique (custom VITAN, which employs the same principles of pixel intensity as OCTOR, requires no manual input other than confirmation of the selected vitreous area)Citation20,Citation23 and one study used manual analysis of OCT images using a subjective observer-based grading system consisting of grades 0–2, where grade 0 was ‘not visible,’ grade 1 was ‘barely visible,’ and grade 2 was ‘clearly visible.Citation24

Index test reliability

Four studies reported index test reliability using varying methodologies. Davis et al. reported an intraclass correlation (ICC) of 0.88 between two observers grading fundus photographs against the 9-point Miami scale.Citation26 Madow et al. reported an inter-observer ICC of 0.87 and an intra-observer ICC of between 0.84 and 0.93 against the Miami scale.Citation18 Keane et al. used Bland–Altman plots to assess interobserver variability and reported a median 95% limits of agreement (LoA) of 0.0353 for all OCTs, 0.0450 in OCTs of uveitic eyes with vitreous haze and 0.0226 for OCTs of healthy eyes or uveitic eyes without vitreous haze. They reported the variance ratio (F statistic) as non-significant between groups, suggesting the measurement variance was similar in eyes with and without vitreous inflammation.Citation19 Mahendradas et al. reported interobserver agreement as Cohen’s kappa >0.7 for all four tested techniques (standard OCT, enhanced vitreous imaging, enhanced depth imaging, and combined depth imaging).Citation24

Correlation between index tests and the clinical reference test: Slit-lamp based clinician grading

Four studies reported correlation between an index test and clinician grading of vitreous inflammation (three studies using OCTCitation19-Citation21 and one study using retinal photography.Citation18) All studies reporting correlation used the NEI VH scale as a comparator. The total number of participants included in these four studies was 307 (430 eyes). Spearman’s r was used by all studies except by Madow et al. to measure the association between index test measurements and the NEI VH scale. The level of correlation between OCT measurements and the NEI VH scale using the semi-automated OCTOR software was 0.53–0.57,Citation19,Citation21 whereas for the fully automated VITAN software correlation was marginally higher at 0.59–0.60.Citation20,Citation28 Both studies by Keane et al., reporting the use of OCTOR and VITAN, used the same retrospective dataset of images. The level of correlation between manual grading of retinal photographs (using the Miami scale) versus clinician examination (using the NEI VH scale) was reported as r = 0.51. The correlation between index tests and the NEI VH scale are shown in . None of the four studies reported confidence intervals for correlation coefficients and those shown in the forest plot were estimated using sample size and correlation coefficient. Passaglia et al. measured agreement between automated fundus photography grading (using the Miami scale) and clinician grading, rather than correlation. They report exact agreement, agreement within one level and agreement within two levels of 0.61, 0.78, and 0.80, respectively, against clinician grading using the NIH scale and 0.67, 0.82, and 0.84, respectively, against the clinician grading using the Miami scale.Citation27

Figure 2. Level of correlation between index tests and clinician grading (SUN/NEI/Nussenblatt vitreous haze scale).

RPE: retinal pigmented epithelium, OCT: optical coherence tomography. *Keane 2015 uses VITAN, an automated version of the previously published OCTOR software. Two variations of the same technique are presented: vitreous:RPE signal ratio and vitreous:RPE textural ratio.** Keane 2014 and Keane 2015 used the same cohort of patients for both studies.
Figure 2. Level of correlation between index tests and clinician grading (SUN/NEI/Nussenblatt vitreous haze scale).

Risk of Bias Assessment

The patient cohorts in the correlation studies were a mixture of uveitis etiologies with a low risk of spectrum bias, except in the retinal photography study by Madow et al., where the risk was not assessable as the underlying etiology was not reported.18 Other than Madow et al., all studies used automated/semi-automated systems to quantify vitreous haze; therefore, it was assumed there was no potential influence from knowledge of the clinician grading. All studies used previously recorded clinician grading (from clinical care or clinical trials data), therefore there was no possibility that the reference test could have been influenced by the index test, which was conducted afterward. Madow et al. did not report whether the fundus photograph readers were blinded to the clinician grading results. Although the time interval between index and reference tests were not explicitly reported by any of the studies, it is presumed that clinician grading and the images acquired were performed on the same visit in all studies, even if image analysis for index tests were done at a later date.

Study heterogeneity

After accounting for overlap between studies in terms of similar imaging techniques and duplicated patient cohorts, there was considerable heterogeneity between the methodology and populations across the included studies. Given this level of heterogeneity, we have not performed any meta-analysis of correlation or test reliability for index tests.

Discussion

This is the first systematic review for identifying noninvasive instrument-based tests for detecting and measuring vitreous inflammation in uveitis. Three technologies were found: ultrasound, retinal photography, and OCT. Ultrasound remains primarily for qualitative assessment of features in the vitreous body and has not been shown to quantify inflammation. Retinal photography and OCT have demonstrated the most potential as methods for quantifying vitreous inflammation through automated and semi-automated means of image analysis. However, only 12 studies have been undertaken and even fewer provided sufficient evidence on test reliability or correlation with clinician grading.

Davis et al. and Madow et al. reported good interobserver reliability (ICC>0.84) and moderate correlation (r = 0.51) of manual grading using retinal photography (assessed using the Miami scale).Citation18,Citation26 This photographic method introduces two advantages beyond the traditional indirect biomicroscopic approach (assessed against the NEI VH scale). Firstly, it captures an adequate view of the fundus and removes the variability introduced by the level of the indirect biomicroscopy skills of the examiner. Secondly, it is based on a 9-point scale rather than the 6-point NEI VH scale, allowing smaller differences to be captured between grades. The automated retinal photography technique applied by Passaglia et al. brings added objectivity beyond the direct biomicroscopic assessment of the NEI VH scale or the original subjective photograph-to-photograph comparison of the Miami grading. On the other hand, the OCT-based technique utilizes signal intensity detected in the vitreous, to derive a measure of light reflectivity as a continuous variable. The ability to detect vitreous reflectivity on a continuous scale means the OCT-based method may potentially offer sensitivity to even smaller, but potentially clinically significant, changes in vitreous inflammation.

Whilst automation of image analysis may improve reliability, we did not find that it consistently improves correlation with clinician grading. The fully automated VITAN OCT algorithm was tested on the same dataset as the semi-automated OCTOR algorithm and showed marginally higher correlation when compared to the NEI VH grade (r = 0.60 versus 0.57).Citation20 Manual grading of retinal photography showed moderate correlation (r = 0.51)Citation18 when compared to the NEI VH scale, similar to the moderate agreement reported for fully automated photographic grading (Cohen’s K = 0.61).Citation27

Strengths and limitations of the review

This review represents the first systematic evaluation of technologies for measuring vitreous inflammation in uveitis. The search strategy was designed to be highly sensitive, using a broad range of databases, including conference proceedings, dissertation databases and the grey literature. This review also has several limitations. An issue in undertaking systematic reviews of correlation between assessment methods is the absence of an adequate specific tool for assessing risk of bias in studies. We have used relevant elements of the QUADAS-2 tool for risk of bias assessment in test accuracy studies for the correlation studies only, where one test was being compared against another. However, this method of assessing risk of bias could not be applied to include studies which only evaluated one test (i.e. for index test reliability). Second, although we included all studies reporting instruments with the potential to detect and measure vitreous inflammation, the data extraction and analysis were focused on test reliability or correlation with the clinical standard. As a result, two studies that provide evidence of the clinical validity and value of new techniques were not discussed in detail.Citation22,Citation23 These include Sreekantam et al.’s study which reported a highly statistically significant reduction of OCT-based vitreous signal (using OCTOR) pre- and post-injection of sub-tenon’s triamcinolone in patients with uveitic macular edema, demonstrating the potential clinical utility of this technique for detecting treatment response and its potentially superior sensitivity for measuring change over the clinician based grading system; however, this study did not include NEI VH scale as a comparator.Citation22 Coric et al. also explored whether a difference in vitreous intensity could be detected in patients with multiple sclerosis versus healthy controls, but did not find a measurable difference; again this study did not include NEI VH scale as a comparator.Citation23 Additional imaging techniques such as ultra-wide field fundus photography (using the Optos ultra-wide field camera) have also been used to detect presence and absence of vitreous haze through manual observation.Citation29 Third, the focus of this review was on correlation with the reference test. Whilst correlation is helpful in early validation, it is limited to demonstrating agreement and non-inferiority to the comparator. From correlation, it is not possible to determine if the index test is more accurate than the reference test. To determine accuracy, a more reliable reference test than the NEI VH scale is required, such as the use of invasive vitreous sampling to determine the level of protein and cellular infiltrates in the vitreous. Due to risks involved, it is unlikely that vitreous sampling will be ethically justifiable in routine practice. In the absence of a reliable reference test, future work could compare the ability of index tests versus clinician grading to detect changes in inflammation, such as the pre- and post-treatment comparison Sreekantam et al. conducted.Citation22 The ability to demonstrate higher sensitivity to small changes as well as stronger association with other inflammatory markers (such as central macular thickness) and visual function, would provide further evidence of accuracy in measuring the true disease state.

Limitations of the evidence

Firstly, due to the small number of included studies and heterogeneity in study design, meta-analyses of correlation or reliability were not possible. Several studies were conducted by the same author groups and presented sequential updates of the same technique using different approaches to image analysis, including automation.Citation20,Citation27 Most studies used retrospectively collected images, with several applying newer analysis techniques to the same image set. Incomplete reporting and varying methodology of the included studies also meant we were unable to pool estimates of correlation between index and reference tests. Secondly, authors sometimes reported correlation coefficients estimated from a mixed cohort of uveitic and healthy eyes. With the exception of Madow et al., where only uveitic eyes were included in the study, all other studies reporting correlation coefficients were a mixture of healthy and uveitic eyes.Citation19Citation21 It was not possible to separate the two cohorts as correlation was reported at an aggregated level in all cases. On the other hand, all studies reporting intra/inter-observer reliability included uveitic eyes only. Thirdly, of those studies that reported NEI VH grading, no patients for OCT and only one patient for retinal photography had grade 4+.Citation18 It could be that in dense vitreous haze, neither OCT nor photography can successfully acquire a usable image and such cases could have been excluded on the basis of poor image quality. However, it is unclear how those index tests performed in the most severe grades of vitreous inflammation.

Clinical relevance and impact

Of the instrument-based tests identified, OCT and retinal photography are presented with the most supporting evidence in this review. Both instruments offer the attractiveness of being technologies already widely available in ophthalmic clinics. Additionally, both techniques can be combined with automated image analysis techniques. OCT additionally offers a measurement which can be continuous and it has also been shown to be sensitive to respond to treatment.Citation22 At this stage there are only a few reports identified for either technology and these reports were mostly retrospective studies with small numbers of subjects. As noted earlier very few patients with severe vitritis are included in these studies, and it is difficult to draw conclusions on the validity of both instruments in the most severe levels of inflammation. It could be argued that, where inflammation is obviously detectable through clinical examination, there is less additional value of quantification by a noninvasive imaging technique. However, clearly, the ideal scenario is to have a technique that is sensitive to changes at both ends of the scale, including detecting worsening or improvement in severe inflammation.

Another major consideration is around generalizability of the study findings in the presence of ocular co-pathology. Of particular concern is media opacity such as cataract, which may cause a similar hazy appearance on fundoscopy and which could degrade image quality on both retinal photography and OCT. Given cataracts are a major complication of chronic intraocular inflammation and ocular steroid therapy, many patients with posterior uveitis have cataracts.Citation30 In the included studies of this review, only Davis et al. reported the exclusion of subjects with cataracts.Citation26 Zarranz-Ventura et al. assessed the use of OCT of patients with uveitis, which also included pseudophakia and patients who had undergone vitrectomy. They demonstrated no observable difference in the measurement for each of these groups compared to phakic and non-vitrectomised eyes, respectively.Citation21

An important area for future work is to evaluate the relationship between instrument-based measures and visual function. Sreekantam et al. reported a correlation coefficient of 0.70 between VIT/RPE-relative intensity and visual acuity, a stronger correlation than was demonstrated when the same OCTOR technique was compared to the NEI VH grading by Keane et al. (r = 0.60).Citation19,Citation22 However, this is not a direct comparison due to different subjects in each study. No other studies explored the association between the index test measurements and visual acuity or any other measure of visual function. Whilst the relationship of visual function to inflammatory activity is complex, often being delayed and indirect, it is worthy of exploration. These tests will be of greatest value if their use enables better control of inflammation, such that vision is maintained in the immediate and long term. It is worth noting that the importance of demonstrating clinical validity through association with visual function was emphasized by regulatory bodies at the American Uveitis Society workshop at the University of California Los Angeles (UCLA) in March 2019 on Objective Measures of Intraocular Inflammation for Use in Clinical Trials.Citation31

If the unreliability of the current reference standard is limiting the evaluation and adoption of emerging techniques, are there any other techniques we should consider as a reference test? As previously discussed, direct sampling of vitreous is unlikely to be ethically justifiable unless it is being done for diagnostic purposes. More invasive tests to quantify vitreous inflammation also exist but are largely unused. Vitreous fluorophotometry is an intravenous fluorescein-based imaging technique which can quantify leakage of dye from the blood-retinal-barrier (BRB) and has been used in the assessment of inflammation of the posterior segment.Citation32 Vitreous fluorophotometry measures leakage by measuring the degree of fluorescence throughout the eye at defined axial points before and after the intravenous injection of fluorescein. It relies on the principle that the amount of leakage is proportional to the degree of BRB breakdown. However, due to its invasive nature, vitreous fluorophotometry is rarely performed and for the most part, has been used as an experimental technique rather than for clinical care.Citation33 Nonetheless, it is worth considering that invasive tests like fluorophotometry may be more direct measures of inflammatory activity and may serve as better reference tests with which to validate newer noninvasive tests. Assuming invasive approaches are not undertaken, evidence supporting new techniques and eventual adoption as a ‘reference standard’ is likely to depend on demonstrating high test reliability, strong association with other evidence of inflammation (such as macular thickness, presence of vitreous cells and other vitreous inflammatory infiltrates, presence of retinal vasculitis and vascular leakage and new active inflammatory lesions), and association with visual function (recognizing that this may not be direct or immediate).

Conclusion

Non-invasive instrument-based tests for measuring vitreous inflammation have the potential to improve reliability and speed compared to clinician grading using indirect ophthalmoscopy. Retinal photography and OCT are two promising technologies with the potential to quantify vitreous inflammation; however, further evidence beyond the proof-of-concept studies identified by this review are required to demonstrate clinical utility. Further evaluation in prospective studies should explore association with other measures of posterior-segment inflammation as well as visual function.

Author contributions

XL, CW, BH: data extraction, manuscript drafting, manuscript reviewing. SB, AA: data extraction, manuscript reviewing. PAK, DJM, AKD project conception, supervision, manuscript reviewing.

Declaration of interest

XL and AKD receive a proportion of their funding from the Wellcome Trust, through a Health Improvement Challenge grant (200141/Z/15/Z). AKD and PAK receive a proportion of their funding from the Department of Health’s NIHR Biomedical Research Centre for Ophthalmology at Moorfields Eye Hospital and UCL Institute of Ophthalmology.