4D measurements of biological and synthetic structures using a dynamic interferometer

Pages S20-S29 | Received 02 Aug 2016, Accepted 20 Feb 2017, Published online: 23 Mar 2017
 

Abstract

Considering the deficiency of time elapsed for phase-stepping interferometric techniques and the need of developing non-contact and on-line measurement with high accuracy, a single-shot phase-shifting triple-interferometer (PSTI) is developed for analysis of characteristics of transparent structures and optical path difference (OPD) measurements. In the proposed PSTI, coupled three interferometers which generate four interference patterns, and a polarizer array is used as phase shifters to produce four spatially separated interferograms with π/2-phase shifts, which are recorded in a single capture by a camera. The configuration of the PSTI allows dynamic measurements (4D measurements) and does not require vibration isolation. We have applied the developed system to examine the size and OPD of cells, and the slope of thin films

Acknowledgements

The author thanks Dra Amalia Martínez García and Dr Gustavo Rodríguez Zurita for their helpful suggestions. The author would like to thank the anonymous reviewers for their valuable comments and suggestions to improve the quality of the paper.