Fatigue failure of the sliding hip screw – clinical and biomechanical analysis

, , , , &
Pages 1364-1372 | Received 04 Jan 2017, Accepted 30 Jul 2017, Published online: 09 Aug 2017
 

Abstract

The study was aimed at the clinical and biomechanical analyses of the sliding hip screw breakage with the use of finite element method. We have identified two patients with the sliding screw breakage. In the first patient, the biomechanical analysis revealed the reduced stress values σHMH not exceeding the yield limit or strength limit of the implant. The yield limit was exceeded in second one. Clinical and biomechanical analyses have demonstrated that adherence to technical requirements of the appropriate osteosynthesis implementation is the principal condition of timely healing since it prevents the material failure.