Implant primary stability with an osteocondensation drilling protocol in different density polyurethane blocks

, , , , , & show all
Pages 14-20 | Received 29 Apr 2020, Accepted 03 Aug 2020, Published online: 25 Aug 2020
 

Abstract

The implant primary stability is a fundamental condition for avoiding implant micro-motions that might result in fibrous encapsulation; its achievement is facilitated by macro- and micro-geometry of the implant, and by the bone density and architecture at the intended implant site. The aim was to evaluate an osteocondensing drilling protocol for dental implant positioning compared to standard protocol on polyurethane block. A total of 40 implants, 20 for each osteocondensing group (Test) and 20 for standard drilling group (Control), were positioned. Insertion torque (IT), removal torque (RT), and Periotest were measured. A IT, RT, and Periotest significant difference was present in favor of the implant osteocondensing protocol. The results suggested that osteocondensing protocol represents a useful technique for implant placement in poor density bone.

Disclosure statement

The authors declare no conflict of interest for this research.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.