Abstract
The properties of BeO thermoluminescent (TL) dosimeters have been investigated in the 1.0–6.0 Gy range. Annealing procedure, fading, calibration and TL dependence on energy and dose were studied along with other relevant physical features. Supralinearity phenomena have been found above 0.5 Gy, but repeated measurements showed a marked stability and reproducibility of the TL response curve vs. dose. The BeO TLDs near tissue equivalence, good sensitivity and integrity to high radiation doses, make them especially suitable for a variety of dosimetric purposes in the radiotherapy field.