Original Articles

Continuous PM-10 Measurements Using the Tapered Element Oscillating Microbalance

&
Pages 1079-1083 | Received 09 Aug 1990, Accepted 08 May 1991, Published online: 06 Mar 2012
 

Abstract

The goal of continuous ambient participate monitoring has been realized through the use of tapered element oscillating microbalance technology. This technology measures particle mass inertially as the mass deposits on a filter. As such, it is a direct mass measurement and eliminates the mass uncertainties inherent in indirect methods represented by optical or beta adsorption instrumentation. A description of the tapered element oscillating microbalance PM-10 measurement technique is given. Tests have been conducted on tapered element oscillating microbalance PM-10 monitors both in the United States and Europe. Test results presented include comparisons to samplers designated as EPA reference methods.