Postcolumn derivatization of amino acids using reaction flow chromatography columns with fluorescence detection: A fast new approach to selective derivatization techniques

, , , , , , , , & show all
 

ABSTRACT

Reaction flow (RF) chromatography with fluorescamine reagent and fluorescence detection (FLD) was used for the analysis of amino acids. The performance of RF chromatography was tested against several optimized conventional postcolumn derivatization (PCD) methods. RF columns achieved greater sensitivity compared to conventional PCD methods, without the need for reaction loops, which resulted in more efficient separations. The RF-PCD method also achieved limits of detection (LOD) from the low picomole to subnanomole range. The calibration data of the RF-PCD technique yielded R2 ≥ 0.99 and % relative standard deviation in peak areas ranging from 0.34% to 5%. Through reaction flow chromatography, multiplexed detection was also achieved allowing the monitoring and analysis of derivatized and nonderivatized flow streams simultaneously.

GRAPHICAL ABSTRACT

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.