Factors Influencing Fast Liquid Chromatography of Proteins and Peptides with Microbore Technology

&
Pages 21-32 | Received 01 May 1992, Accepted 05 Jun 1992, Published online: 23 Sep 2006
 

Abstract

Microbore columns (1 and 2. 1 mm I. D.) were investigated for their utility in fast liquid chromatography of proteins and peptides. Using a series of SynChropak large pore reversed phase columns, the effects of flowrate and gradient on resolution of standard protein and peptide mixtures were determined. Excellent resolution could be achieved on 2.1 mm I. D. columns with analysis times under five minutes.