240
Views
8
CrossRef citations to date
0
Altmetric
Original Articles

Fabrication of an improved amperometric creatinine biosensor based on enzymes nanoparticles bound to Au electrode

, , &
Pages 739-749 | Received 09 Aug 2019, Accepted 04 Oct 2019, Published online: 24 Oct 2019
 

Abstract

An improved amperometric creatinine biosensor was fabricated that dependent on covalent immobilisation of nanoparticles of creatininase (CANPs), creatinase (CINPs) and sarcosine oxidase (SOxNPs) onto gold electrode (AuE). The CANPs/CINPs/SOxNPs/AuE was characterised by scanning electron microscopy and cyclic voltammetry at various stages. The working electrode exhibited optimal response within 2 s at a potential of 0.6 V, against Ag/AgCl, pH 6.5 and 30 °C. A linear relationship was observed between creatinine concentration range, 0.1–200μM and biosensor response i.e. current in mA, under optimum conditions. Biosensor offered a low detection limit of 0.1 μM with long storage stability. Analytical recoveries of added creatinine in blood sera at 0.5 mM and at 1.0 mM concentrations, were 92.0% and 79.20% respectively. The precision i.e. within and between-batch coefficients of variation were 2.04% and 3.06% respectively. There was a good correlation (R2 = 0.99) between level of creatinine in sera, as calculated by the colorimetric method and present electrode. The CANPs/CINPs/SOxNPs/Au electrode was reused 200 times during the period of 180 days, with just 10% loss in its initial activity, while being stored at 4 °C, when not in use.

    Highlights

  1. Prepared and characterised creatininase (CA), creatinase (CI) sarcosine oxidase (SOx) nanoparticles and immobilised them onto gold electrode (AuE) for fabrication of an improved amperometric creatinine biosensor.

  2. The biosensor displayed a limit of detection (LOD) of 0.1 μM with a linear working range of 0.1 μM–200 μM.

  3. The biosensor was evaluated and applied to measure elevated creatinine levels in sera from whom suffering from kidney and muscular disorders.

  4. The working electrode retained 90% of its initial activity, while being stored dry at 4 ˚C for 180 days.

Correction Statement

This article has been republished with minor change. This change do not impact the academic content of the article.

Disclosure statement

The authors declare that there is no conflict of interests regarding the publication of this paper.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.