Objectives: Cochlear implantation (CI) is considered an effective and relatively safe procedure for patients with severe-profound hearing loss. However, severe complications are reported in several studies. The purpose of this study was to report the frequency and management of major complications following CI surgery at the National CI Programme (NCIP) in Ireland.

Methods: Major complications were defined according to the classification of Hansen et al. 2010. The medical records of 1017 patients undergoing CI between 1995 and 2016 were analyzed retrospectively for major complications. In addition, radiological and intraoperative findings as well as therapeutic management of all patients with a major complication were reviewed.

Results: Altogether, 1017 patients underwent 1266 CI surgeries. The median follow-up of all CI surgeries was 44 months. The total number of major complications identified was 21 which corresponds to an overall rate of 1.7%. The majority of major complication (71%) occurred at least one week after surgery. The most common major complications were internal receiver/stimulator protrusion and migration (6/21), followed by migration of the electrode array (4/21) and recurrent otitis media requiring re-implantation (4/21). All patients with major complications required additional surgery, with reimplantation necessary in 19 patients (90%).

Discussion: CI surgery is a safe surgical procedure for hearing rehabilitation associated with a low rate of severe complications. However, major complications can occur many years after surgery, making revision surgery necessary.

Conclusion: Long-term follow-up is necessary for the early identification of complications to facilitate appropriate care.

Acknowledgments

The authors would like to thank the whole team of the National Cochlear Implant Programme for their supportive work in data collection and constructive discussions.

Disclaimer statement

Contributors None.

Funding None.

Conflict of interest None.

Ethics approval None.

ORCID

Cristina Simoes-Franklin http://orcid.org/0000-0002-4099-2708

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.