COMPARISON OF IN-HOUSE DEVELOPED ELISA WITH HPLC FOR MEASUREMENT OF HBA1C

, &
Pages 305-312 | Published online: 08 May 2013
 

Abstract

In-house developed ELISA for HbA1c was evaluated in comparison with Bio-Rad D-10 Dual Program (HPLC). % HbA1c was estimated using said methods in 40 blood samples. Intra and inter assay coefficient of variation (CV) were in acceptable range i.e., <3%. HbA1c results obtained by two methods showed good correlation, (y = 0.8942x + 0.7229 R2 = 0.877). Sensitivity, specificity, positive, and negative predictive value for in-house ELISA classifying patients as controlled and uncontrolled diabetes against HPLC were recorded as 87%, 94.1%, 95.0%, and 93.8%, respectively, while accuracy of the test was 90.0%. In-house ELISA for HbA1c is not only accurate, cost effective and simple but also precise comparable to HPLC in deciding glycemic status of patients.

ACKNOWLEDGMENT

The authors gratefully acknowledge the support of the late Dr. Raymond Edwards, from NETRIA UK, who provided us with the reagents and technical advice for developing this assay. We are also thankful to Simon Packer (SCIPAC Ltd.) for supplying HbA1c controls free of cost.

Notes

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.