Effect of Upper Strap Downward Displacement on N95 Filtering Facepiece Respirator Fit Factors: A Pilot Study

, , , &
 

Abstract

Fifteen subjects underwent three replicates of quantitative respirator fit-testing with N95 filtering facepiece respirators that were donned with the upper strap high on the occiput, as per the manufacturers’ donning instructions. Each fit-test was immediately followed by repeat fit-testing with the upper strap downwardly displaced to the level of the ear sulcus to determine any change in fit factors that might occur with upper strap downward slippage. A total of 35/45 (78%) initial fit-tests had a passing score (fit factor ≥100) with the top strap high on the occiput and 33/35 (94%) of these passed subsequent fit-testing after the top strap was displaced downward to the ear sulcus. Geometric mean fit factors for the initial passed fit-tests, and following downward strap displacement, were 217±1.6 and 207±1.9, respectively (p = 0.64). Downward displacement of the top strap did not significantly impact fit factors of N95 FFRs that had previously passed fit-testing.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.