ABSTRACT
Background: Current advances in refractive surgery require high accuracy of corneal measurements. We aim to evaluate the repeatability of topographic measurements using an integrated Placido disk - OCT device.
Methods: Thirty right eyes of thirty healthy and young subjects were included. The topographic parameters of simulated keratometry, asphericity, minimum pachymetry, white-to-white distance, spherical aberration and corneal elevation were studied. Three measurements were taken by the same observer using the Visante omni device. The coefficient of repeatability (CoR), coefficient of variation (CoV) and intraclass correlation coefficient (ICC) were calculated to assess the repeatability. The Bland-Altman graphical method was also applied.
Results: No significant differences between repeated measurements were found. The mean difference for simulated keratometry was between 0.01 and 0.02 D. For asphericity was −0.01 ± 0.02, and for minimum pachymetry, spherical aberration and WTW was equal or less than 0.02 mm. The mean differences for corneal elevation, in millimeters and diopters, were between 0.00 and 0.01. Overall, most topographic parameters had a CoV less than 0.4%, a CoR less than 0.3 and an ICC higher than 0.98.
Conclusions: The Visante omni device provides good repeatability for topographical measurements of simulated keratometry, asphericity, pachymetry, spherical aberration and corneal elevation.
Declaration of interest
The authors have no relevant affiliations or financial involvement with any organization or entity with a financial interest in or financial conflict with the subject matter or materials discussed in the manuscript. This includes employment, consultancies, honoraria, stock ownership or options, expert testimony, grants or patents received or pending, or royalties.