ABSTRACT
Background: After encouraging results with the Edwards Sapien and XT valves, this study aimed to review procedural data and early outcomes for the Sapien 3 valves for transcatheter pulmonary valve replacement (TPVR).
Methods: We performed a multicenter, retrospective analysis of cases who underwent a Sapien 3 TPVR between 2015 and 2017 in 7 centers in Germany with a follow-up of up to 2 years.
Results: 56 patients could be enrolled (weight 58,5 ± 25,0 kg; 53% Tetralogy of Fallot, 45% native RVOT). Most procedures were two-stage procedures (82,1%) with 100% prestenting. Valve sizes were 20 mm (n = 1), 23 mm (n = 15), 26 mm (n = 27), 29 mm (n = 13). Procedural success rate was 96.4%. Two patients underwent surgical valve implantation after balloon rupture during TPVR. Follow-up data were available up to 24-month post TPVR. The rate of patients with ? moderate and severe pulmonary regurgitation decreased to 0% after TPVR, peak systolic gradient decreased from 24,2 (SD±20,9) mmHg to 7,1 mmHg (SD±5,0). There were no endocarditis, severe tricuspid valve impairment or stent fractures.
Conclusions: With the Edwards Sapien 3 valve, the patient pool for TPVR can be substantially extended. Continued data collection is necessary to verify long-term results.
Declaration of interest
NA Haas is a proctor for Edwards Sapien valves. The authors have no other relevant affiliations or financial involvement with any organization or entity with a financial interest in or financial conflict with the subject matter or materials discussed in the manuscript apart from those disclosed.
Reviewer disclosures
Peer reviewers on this manuscript have no relevant financial or other relationships to disclose.