Nocturnal Transcutaneous Blood Gas Measurements in a Pediatric Neurologic Population: A Quality Assessment

, &
Pages 303-310 | Received 18 Aug 2020, Accepted 22 Dec 2020, Published online: 11 Jan 2021
 

ABSTRACT

Objective: To assess the quality of SpO2 and PCO2 recordings via transcutaneous monitoring in children with neurological conditions.

Methods: Overnight transcutaneous SpO2 and PCO2 were analyzed. The presence of drift and drift correction was noted, and the rate of disrupted recordings scored (0: absence, 1; presence). The quality of recordings was also scored (0, 1, 2 for poor, medium, and high).

Results: A total of 228 recordings from 64 children aged 9.7 ± 6 years were analyzed of which 42 used positive pressure respiratory support. The mean quality of the recordings was scored as 1.27 (0–2). PCO2 drift, drift correction, and disrupted recordings were present in 25%, 58%, and 26% of recordings, respectively. Satisfactory clinical decisions were taken in 91% of cases.

Conclusion: The quality of transcutaneous sensor recordings was acceptable and clinical findings were deemed as satisfactory in the large majority of cases. Correction of PCO2 drift was challenging.

Highlights

Disclosure Statement

The authors declare to have no conflict of interest.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.