Influence of electro-polishing characteristics on needle-tissue interaction forces

ORCID Icon &
 

Abstract

The electro-polishing process was applied to stainless steel 304 with the aim of improving its surface texture. The electrolyte was a mixture of H2O, H2SO4 (98%) and H3PO4 (85%) at a volume ratio of 2:3:5. Gravimetric analysis and stylus profilometry were used to analyze the samples. It was observed that the material removed during the electro-polishing process could be controlled to an error limit of 6%. The surfaces obtained showed improved cleanness and reduced surface roughness (Ra). Hypodermic needles were then electro-polished using the same parameters with the aim of reducing forces during needle insertion into tissue. The needle tips were analyzed using a SEM. The needles were tested using silicone phantoms as a tissue substitute. The forces during insertion were less for all the electro-polished needles, with the most consistent results obtained at a current density of 0.66 A/cm2. A theoretical model of the needle insertion forces at different insertion depths was developed.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.