Point-of-care estimation of haemoglobin concentration in all age groups in clinical practice

, MBBCh, FCPath(Haem), MMed(Haem) (Consultant) & , MBBCh, FCPath(Haem), MMed(Haem), BSc, Cert Clin Haem(SA) , Principal Haematologist
Pages 271-274 | Received 05 Sep 2010, Accepted 20 Dec 2010, Published online: 15 Aug 2014
 

Abstract

Background: The measurement of haemoglobin (Hb) concentration provides a reliable, primary screening test for the presence and severity of anaemia. The HemoCue® is a point-of-care test for Hb measurement. The introduction of point-of-care testing in hospitals and general practice has improved patient care and accessibility. This study was performed to evaluate the utility of point-of-care Hb measurement with the HemoCue® device for the diagnosis of anaemia.

Method: In this prospective study, we compared the analytical performance of the HemoCue® to the Coulter® LH 750 automated haematology analyser with regard to accuracy, precision and linearity in the measurement of Hb in adult and paediatric patient samples, referred for routine laboratory testing.

Results: Samples from 100 patients were analysed with both instruments, and the results were compared using standard scatter and difference plots. The mean Hb value of the HemoCue® (11.3 g/dl; range 4.6–16.7) was comparable to the Coulter® LH 750 (11.3 g/dl; range 4.7–17.2). The Bland-Altman difference plot revealed good correlation. Bias between the two methods was small, and the imprecision was within acceptable limits. Hb measurement was linear in the range of 4.8–20 g/dl.

Conclusion: In all age groups, the HemoCue® point-of-care device can be used to provide accurate and reliable Hb measurements with a smaller sample volume, improved turnaround time, and long-term cost saving.