57
Views
3
CrossRef citations to date
0
Altmetric
Original Articles

Determination of Scutellarin in Scutellaria barbata Extract by Liquid Chromatography–Electrochemical Detection

, , , &
Pages 2133-2140 | Received 30 Nov 2002, Accepted 12 Feb 2003, Published online: 16 Aug 2006
 

Abstract

A rapid and simple method was developed for the determination of scutellarin in Scutellaria barbata plant extracts by reversed‐phase liquid chromatography with an electrochemical detector (ECD). The potential electrode voltage was +0.7 V for the ECD. The mobile phase consisted of methanol:water:glacial acetic acid (30:68:2, v/v/v) (pH 2.7). The linear calibration ranged from 2.3 to 150.0 µg mL−1 for scutellarin. The detection limit was 0.2 µg mL−1 given a signal‐to‐noise ratio (S/N) of 3:1 on the second detection fullscale (500 nA). The six different extracts of S. barbata were determined using different extraction methods and extraction solvents. The contents of scutellarin in the extracts were determined within 25 min.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.